7
Features of the Product
Maintenance
If the monitor has not been used recently, be sure to verify that it operates correctly before using it.
(Page 16)
Durable period
6 years, provided that the appropriate maintenance has been done. (Self-certification through
OMRON HEALHCARE's own data)
Measures to be taken if a problem or accident occurs
If an equipment error (E9) occurs, take the following measures promptly.
1. Remove the arm from the cuff.
2. Turn off the power and unplug the monitor.
3. Place an "Out of Service" note on the monitor so nobody will use it.
4. Please contact your local authorized OMRON dealer/supplier.
Features of the Product
• Supports accurate measurement and can easily be used by a single person.
Includes a "Correct Measurement Support" function that enables even beginners to assume the
correct body position.
This prevents measurement errors due to incorrect body position.
• When the elbow rest detects the elbow, the elbow detection lamp lights up.
• The moveable arm cuff allows even tall people to assume the correct body position.
• Measurement is possible over a wide range of arm circumferences (17 to 42 cm).
• Left or right arm can be used for measurement.
• Oscillometric measurement.
The monitor can detect low and soft Korotkoff's sound, and be used in a moderately loud
environment.
During cuff depressurization, the monitor determines blood pressure by detecting the pressure
pulse wave that occurs simultaneously with the heart beat.
• Measurement results can be announced.
• Customised printed results.
Logo/image can be uploaded and printed on the results print out.
• The unit and cuff cover are antibacterial.