34
Manufacture's Declaration
Electromagnetic Immunity (IEC60601-1-2)
Immunity
Test
IEC60601-1-2
Test Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±2,4,6 kV contact
±2,4,8 kV air
±2,4,6 kV contact
±2,4,8 kV air
Floor should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electric fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/
output lines
±2 kV for power
supply lines
±1 kV for input/
output lines
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Surge
IEC 61000-4-5
±1 kV normal
mode
±2 kV common
mode
±1 kV normal
mode
±2 kV common
mode
Mains power quality should be that of a
typical commercial and/or hospital
environment.
Voltage dips,
short interruptions
and voltage
variations on
power supply
IEC 61000-4-11
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
Mains power quality should be that of a
typical commercial and/or hospital
environment.
If the user of the HBP-9020 requires
continued operation during power mains
interruption, it is recommended that the
HBP-9020 be powered from an
uninterruptible power supply.
40 % UT
(60 % dip in UT)
for 5 cycles
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(95 % dip in UT)
for 5 sec.
<5 % UT
(95 % dip in UT)
for 5 sec.
Power frequency
(50/ 60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or hospital
environment.
Note: UT is the AC mains voltage prior to application of the test level.
HBP-9020.fm 34 ページ 2010年11月29日 月曜日 午前11時8分
Содержание HBP-9020
Страница 2: ......
Страница 24: ...22 Printing the Measurement Data Print pattern B Print pattern C 3 2 1 4 5 3 2 1...
Страница 39: ......