6
Notes on Safety
General Advice
• Do not place anything on this monitor.
• Be sure to inspect the monitor before starting and perform regular maintenance inspections to ensure safe and
proper use of the monitor.
• If a measurement error occurs or a result is questionable, use auscultation or other method to check.
- Measured values may be indicated erroneously or measurement may not be possible for a person who has
significant short-term fluctuation of blood pressure such as an irregular pulse.
- For people who have low peripheral blood flow, significantly low blood pressure, or significantly low body
temperature, measurement may not be possible (an error will occur) or the measurement result may not be
reliable due to insufficient blood flow in the measurement location.
- Measurement reliability will decrease if the user moves or speaks during measurement, or the body position is
not correct.
- Measurement reliability will decrease if the arm is compressed because a garment with thick sleeves is worn or
the sleeves are rolled up.
- Measurement reliability will decrease after bathing or showering, drinking an alcoholic beverage, smoking
tobacco, exercising, or eating.
• Do not install the monitor in a location that is exposed to direct sunlight for an extended period of time.
• Do not store this monitor in the following locations.
- A location where water may splash onto the monitor
- A place where the device receives the direct sunlight for a long time
- Place filled with plenty of dusts, salt, and sulfur in the air
- Slanted place or the place that receives vibration or shock (including while carrying the monitor)
- In a location where chemicals are stored or corrosive gas is generated
- In a location that is not within the suitable temperature and humidity range (ambient temperature -20 to 60°C,
humidity 10 to 95% RH)
• Do not block or place objects in the printer paper outlet.
- This will cause paper jams.
• Do not attempt to pull out the printer paper until it is automatically cut.
• Do not subject the monitor to shock or drop the monitor.
• Do not use a solvent, such as thinner or benzene, for cleaning this device.
- It causes damage to its exterior.
• Read the instruction manual of the other devices to be used at the same time with this unit to understand and be
aware of the interaction between the devices.
• Do not measure without inserting the arm in the cuff.
HBP-9020.fm 6 ページ 2010年7月9日 金曜日 午前8時59分
Содержание HBP-9020
Страница 2: ......
Страница 24: ...22 Printing the Measurement Data Print pattern B Print pattern C 3 2 1 4 5 3 2 1...
Страница 39: ......