19
D6F-PH MEMS Differential pressure Sensor User’s Manual (A288)
8-3.
Example of Sensing data
The below tables provide a few basic examples of the data that is read in Flow Sensor
Registers when the device is subject to given flow and temperature. The values listed in the
tables are given under the hypothesis of perfect device calibration (i.e. no offset, no gain error,
etc).
Table14. Temperature Data registers content vs. Temperature value
Address of Registers
Value of Registers
Converted
Temperature values
TMP_H
TMP_L
D061h
D062h
HEX
DEC
2Bh
8Dh
2B8Dh
11149
25.0 degC
2Eh
FFh
2EFFh
12031
48.6 degC
26h
BBh
26BBh
9915