At this time you should be able to access the main menu. This pin code resets the system and you should
re-program your pin code again. Do not re-enter the system with the pin code of “1950”. It may cause all
data to be cleared.
If the problem persists, contact OmniGuide for assistance
13.3.
No effect on tissue
1.
Make sure the fiber is not clogged. Clean in sterile solution if necessary. Replace fiber if necessary.
2.
Increase the power of the system
3.
Is the “Ready” button pressed?
4.
Increase the pulse length
5.
Do you hear the trigger beep when the foot pedal is pressed?
If the problem persists, contact OmniGuide for assistance
13.4.
Low power or no power is observed out of the OmniGuide Fiber:
1.
Verify that the Laser is on and that it is set to “Ready”.
2.
Verify that gas of the proper type and pressure is being supplied to the gas hose of the Adapter. The
proper gas type and pressure is shown on the insert provided with each fiber.
3.
Verify that the Adapter is securely mounted on the Laser. If the Adapter exhibits any up-and-down,
rocking, or rotational movement relative to the Laser, discontinue use and contact OmniGuide for
assistance.
4.
Set the Laser power to 4W in CW mode. Keep the cannula of the handpiece (if present) and the fiber
as straight as possible and operate Laser for up to 5 seconds while the fiber’s distal-end is 3mm from
a moist tongue depressor. Use aseptic technique. If no visible burn spot appears, replace the
OmniGuide Fiber and repeat this test.
If the problem persists, contact OmniGuide for assistance
13.5.
System too hot
1.
Is the filter system cleaned?
2.
Is no restriction in the air flow at the bottom of the system?
3.
Is the room temperature below 28° C (82F)?
4.
Have you treated patients with full power more than 30 minutes?
5.
Is the system close to any heating?
If the problem persists, contact OmniGuide for assistance.
Warranty
14.1.
Warranty on parts
OmniGuide warrants that the parts are free from defects in materials and workmanship. For specific
warranty terms and conditions for your FELS-25A Laser system refer to your sales contract. The FELS-
25A is warranted for parts for a period of 12 months. Warranty begins from date of shipment. Any
specific terms with your distributor are part of the purchase agreement between you and your
distributor.
Содержание FELS-25A
Страница 10: ...L 1633 Rev 3 OmniGuide Inc Page 10 1 Warning Aperture 2 Laser Stop 3 Laser Warning 4 Laser Danger...
Страница 11: ...L 1633 Rev 3 OmniGuide Inc Page 11 5 Laser Compliance 6 ETL 7 Modification Label 8 Product Identification...
Страница 23: ...L 1633 Rev 3 OmniGuide Inc Page 23 Figure 5 7 Dimensions of the FEL 25A Laser...
Страница 24: ...L 1633 Rev 3 OmniGuide Inc Page 24 Figure 5 8 Dimensions for the Fiber Enabled Laser on OG designed Cart...