12.2.
Alignment:
The OmniGuide optical adapter is to have its alignment confirmed and maximized to
within specification. In addition the laser beam should be inspected to ensure the beam profile (laser
modes) and wavelength shifts are acceptable for optimal performance.
12.3.
Calibration:
The power outputs of the laser are to be measured at several power levels from 1-20 for
high level software versions 1.3 and prior and from 1-30 for high level software versions 1.4 and
higher. Measured using continuous, and super pulse wave forms. The power should be within 10%
of the laser power setting measured after the final turning mirror and before the focusing lens.
12.4.
Gas Management:
The gas management assembly is to be measured for proper output pressure and
inspected to ensure the tank levels are properly displayed. It is also to be inspected for any leaks or
damaged connections.
12.5.
Parts:
The FELS-25A laser, cart, and accessories are to be inspected for any loose, damaged, or
missing parts. The following parts are to be replaced annually
12.5.1.
Fan filter element (located in the filter cassette shown in figure 3-2)
12.5.2.
¼” o-rings for the high pressure hose from the gas management assembly to the rear of
the laser
12.5.3.
1/8” o-rings for the low pressure hose from the front of the laser to the adapter.
12.5.4.
Yoke washers for the CGA-930 fitting that connects the inlet hoses for the gas
management assembly to the tanks
12.6.
Example Report:
Please see section 20 of this document for an example service report covering all
required maintenance parameters.
12.7.
Additional Information:
for additional information on the Yearly Preventive Maintenance please
contact OmniGuide Service at 1-888-666-4484 ext. 3
Troubleshooting
13.1.
System does not start
1.
Check if electrical power cord is attached to the main supply
2.
Check if electrical power cord is plugged into the Laser
3.
Check if mains switch at rear is in “on“ position
4.
Check if emergency switch is pulled out all the way
5.
Check if interlock connector is in place at rear of Laser
6.
Check if there is the right voltage on the mains socket.
7.
Check fuses at the mains connector at rear of the Laser. If the fuses are blown, exchange fuses with
spare fuses supplied with the system.
8.
Door interlocked and door is open.
If the problem persists, contact OmniGuide service for assistance (1-888-666-4484 x3)
13.2.
Password does not work
Retype the pin code (password). If you cannot get into the main menu restart the system and retry. If
you have no success entering the main menu press following numbers:
1950
Содержание FELS-25A
Страница 10: ...L 1633 Rev 3 OmniGuide Inc Page 10 1 Warning Aperture 2 Laser Stop 3 Laser Warning 4 Laser Danger...
Страница 11: ...L 1633 Rev 3 OmniGuide Inc Page 11 5 Laser Compliance 6 ETL 7 Modification Label 8 Product Identification...
Страница 23: ...L 1633 Rev 3 OmniGuide Inc Page 23 Figure 5 7 Dimensions of the FEL 25A Laser...
Страница 24: ...L 1633 Rev 3 OmniGuide Inc Page 24 Figure 5 8 Dimensions for the Fiber Enabled Laser on OG designed Cart...