Laser head and consequently a drop of power. Operating with dust, water or dirt in the environment around the
Laser or without the filter could result in damage to the Laser.
__________________________________________________
Never operate the Laser without the air filter, do not use any other filter than the one
provided by OmniGuide.
____________________________________________________________
9.2.
Electrical Power Connection
The FELS-25A requires AC voltage 90-260 Volts , 50/60 Hz.
The mains cable should be rated above 12A continuous AC current Use main cable supplied with the
system.
Make sure that the mains ground connection is made. . The mains cable is connected to the rear of the
Laser into the socket with the following characteristics:
Input voltage: 90-260 V, 50/ 60 Hertz
Ground connection required
Leakage current trip: less 300 µA
Input current fuses 90 V / 12A, 115 V / 10A, 260 V / 5A
The mains socket should be properly fixed to the wall; it is not allowed to use an extension cable
between the Laser and the wall outlet.
9.2.1.
Fuse Ratings for the Laser
Different input voltages require a different fuse rating.
230 V 5A
115 V 10A
90 V 12A
9.3.
Compressed Gas Requirement
The OmniGuide Laser system is designed to be used with medical-grade Helium gas, delivered to the
fiber at a pressure that depends on the type of fiber and the application. Please see fiber IFU for details.
The gas flow is established by connecting the gas hose (supplied with the system) to a Helium source
(provided by the user). The gas hose has a 1/8” quick-disconnect fitting to facilitate this attachment. The
helium source is then connected to the gas port at the back of the laser. The gas outlet on the front of the
laser has a 1/8 “ quick disconnect port. The gas hose supplied with the system connects the gas outlet at
the front of the laser to the fiber adapter. A Gas Filter Unit (P/N ACC-GFU-100-1) between the gas hose
and the adapter is required for sterile procedures.
9.4.
Door Interlock
The Laser is installed in a room which is clearly marked as Laser room and could be interlocked through
the Laser console. To allow the door to be wired into the Laser, the Laser console has an interlock
connector at the rear of the housing
.
Содержание FELS-25A
Страница 10: ...L 1633 Rev 3 OmniGuide Inc Page 10 1 Warning Aperture 2 Laser Stop 3 Laser Warning 4 Laser Danger...
Страница 11: ...L 1633 Rev 3 OmniGuide Inc Page 11 5 Laser Compliance 6 ETL 7 Modification Label 8 Product Identification...
Страница 23: ...L 1633 Rev 3 OmniGuide Inc Page 23 Figure 5 7 Dimensions of the FEL 25A Laser...
Страница 24: ...L 1633 Rev 3 OmniGuide Inc Page 24 Figure 5 8 Dimensions for the Fiber Enabled Laser on OG designed Cart...