Omnicomm Configurator (Android):
If during gauging the Omnicomm Configurator (Android) does not display all the
connected Omnicomm LLS sensors, select “Refresh the list of devices” in the
menu.
4. Set the flow interval in liters
Perform filling by a measuring container or under control of a fluid flowmeter with
predefined interval. The container should have metrological calibration test
certificate.
5. Press the “Start/continue gauging” button
6. Fill in the fuel volume equal to the flow interval
7. Press “Add line”
The filling volume equal to the predefined flow interval will be displayed in the
“Liters” column.
The value equal to the filling volume will be displayed in the “Sensor” column.
8. Press “Add line”
9. Repeat items 6, 7 and 8 according to the number of control points. The minimum
recommended number of control points – 20
Gauging
Omnicomm LLS 4 Fuel Level Sensors
15