Omnicomm Configurator (Android):
“Empty/Full” Calibration
Perform calibration in the fuel, in which the fuel level sensor Omnicomm LLS will
operate.
1. Fill the measuring container with fuel
2. Immerse the Omnicomm LLS sensor in the fuel to the full length of the measuring
piece
3. Wait for green indicator “Level Stabilized” to appear.
In the “Settings” tab, “Empty/Full Calibration” section, press “Full” button, then the
value corresponding to the full tank will be set
4. Take the Omnicomm LLS sensor out of the container and allow the fuel to flow
down the measuring piece for 1 minute. In the “Empty/Full Calibration” section,
press “Empty” button, then the value corresponding to the empty tank will be set
5. Press the “Write to Device” button
Setting
Omnicomm LLS 4 Fuel Level Sensors
10