Calibration
Calibration of the fuel tank is necessary to verify the conformity of the digital code issued by
the Omnicomm LLS 4 sensor to the fuel volume in a particular fuel tank.
Calibration of the fuel tank is performed by filling up the tank – from empty to full, with a
certain filling interval, and recording the Omnicomm LLS sensor readings in the calibration
table.
A container may be calibrated by draining.
The calibration of a container with multiple Omnicomm LLS 30160 sensors is performed
similarly to the calibration with one sensor. Before the calibration process, add the neces-
sary quantity of sensors and specify the network addresses. Calibration is performed for all
sensors at the same time. Connect multiple sensors to a PC or tablet using a KTZ splitter.
Calibration of the tank with one Omnicomm LLS sensor:
1. Empty the fuel tank
2. Connect the sensor to a PC or tablet according to the diagram in the
3. Run the Omnicomm Configurator program on your PC or tablet. Select “Tank calibration”
operating mode
Calibration
Omnicomm LLS 30160, LLS-AF 4 Fuel Level Sensors
17