1-14
Electronics Tray
Electronics Tray
OmniTT/AnesthesiaTT Installation and Service Guide/67-2024 Rev D
© 2010 Omnicell, Inc.
6. Connect the touch screen data cable to J2 on the motherboard:
a. Connect the extension cable provided in the kit to the touch screen data cable (red wire to
red wire).
b. Wrap the joined connectors with electrical tape to ensure that the cables do not become
disconnected.
c. Connect the touch screen data cable extension to J2 on the motherboard.
Figure 1-23.
Connect the Touch Screen Data Cable to J2 of the Motherboard
7. Connect the inverter cable (backlight) to J32 on the motherboard.
Figure 1-24.
Connect the Backlight Cable to J32 of the Motherboard