Generator Probe Compatibility

Size

Processing 

Range

Length

Type

Resulting 

Particle Size

Order No.

10 mm

1.5 mL - 100 mL

95 mm

Saw Tooth

Fine

15051

195 mm

Saw Tooth

Fine

150510

20 mm

50 mL - 2 L

100 mm

Saw Tooth

Fine 

15201

Coarse

15201W

145 mm

Saw Tooth

Fine

15401

Coarse

15401W

195 mm

Saw Tooth

Fine

15020

Coarse

15020W

30 mm

75 mL - 10 L

195 mm

Flat Bottom

Fine

150-30NA-195

Flat Bottom 

Wide Window

Coarse

150-30WA-195

45 mm 

(Macro Only)

200 mL - 25 L

195 mm

Flat Bottom

Fine

150-45NA-195

55 mm 

(Macro Only

300 mL - 30 L

195 mm

Flat Bottom

Fine

150-55NA-195

Generator Probe Installation

All generator probes (sold separately) are easily installed to the motor unit by following these steps:
1.  Screw the generator probe adapter (P/N 15001) into the coupling adapter
2.  Screw the generator probe into the generator probe adapter until the O-ring is completely inside 

the generator probe adapter.

3.  If the generator probe does not thread fully into the generator probe adapter, then unscrew the 

generator probe from the generator probe adapter. Slightly rotate the rotor shaft collar on the 

generator probe. Then return to step 2.

4.  Tighten the generator probe by hand with moderate force.
5.  Remove the blue protective cap from the tip of the generator probe and the homogenizer is ready 

to operate.

NOTE:  The tip of the generator probe, especially on the saw tooth generator probes, is sharp. For safety 

purposes it is advisable that the protective cap be replaced on the generator probe when not in use. 
NOTE: The bottom of the generator probe is extremely fragile and care should be taken to protect it. 

Replace the blue protective cap on the end of the generator probe when the generator probe is not 

being used.

WARNING

: Never operate the motor with the generator probe (or sealed chamber assembly) partially 

threaded onto the motor. To prevent damage to the motor and attachments be sure that the generator 

probe (or sealed chamber assembly) is fully threaded into the coupling adapter. Failure to do so will 
result in damage to the coupling system.

Содержание Mixer

Страница 1: ...Omni Mixer Macro User Manual OMNI MIXER HOMOGENIZER www omni inc com...

Страница 2: ...SAVE THIS OWNER S MANUAL The Omni Mixer and Omni Macro has been engineered for economical functionality as well as safety however basic safety precautions and common sense must always be demonstrated...

Страница 3: ...Instruction Manual 17258 17258 1 Overview 1 2 3 4 5 1 Support Post 2 Coupling Adapter 3 Support Base 4 Clamping Knob Assembly 5 Motor Support Housing Description P N QTY Gen Probe Multi Tool 00 402 2...

Страница 4: ...robes can be used in open containers or in sealed chamber assemblies Sample process ing efficiency is effected by Amount of material processed vs size and speed of the generator probe Container geomet...

Страница 5: ...ot thread fully into the generator probe adapter then unscrew the generator probe from the generator probe adapter Slightly rotate the rotor shaft collar on the generator probe Then return to step 2 4...

Страница 6: ...eleased Determining Motor Speed To determine motor speed during operation a tachometer not in cluded may be inserted into the Omni Mixer or Omni Macro through the top of the motor as shown by removing...

Страница 7: ...g wear or part failure NOTE Inspect glass or plastic chambers for any scratches or cracks and do not use if any are found Eye protection and utilization of a secondary container is recommended to prev...

Страница 8: ...n the assembly securely 2 Attach the chamber assembly to the motor by screw ing the chamber cover into the coupling adapter 3 If the chamber assembly does not thread fully into the coupling adapter th...

Страница 9: ...commutator and should be re placed check both brushes since both brushes generally do not wear at the same rate If one brush is worn replace both brushes Make sure that the brushes are installed prop...

Страница 10: ...d turned on but is not functioning Check power cord connectors Check wall socket for power to the outlet Motor is turned ON and makes a buzz ing sound but is not working Brushes may need to be replace...

Страница 11: ...This page left blank intentionally...

Страница 12: ...PN 17258 Rev B 935C Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...

Отзывы: