Troubleshooting 

Do not attempt to service the Omni Mixer or Macro in a manner other than those dis-

cussed in this manual.  For any issue that is unsuccessfully corrected using this guide, 

please contact your authorized dealer or call technical assistance at Omni International 

at 1-800-776-4431 or 770-421-0058.

PROBLEM

CORRECTIVE ACTION

The Omni Mixer/Macro is plugged in and 

turned on but is not functioning.

- Check power cord connectors

- Check wall socket for power to the outlet

Motor is turned ON and makes a “buzz-

ing” sound, but is not working.

- Brushes may need to be replaced.

  Contact Omni International or distributor.

Motor unit operating speed declines, 

stalls intermittently, or stops completely

- Brushes may need to be replaced.

  Contact Omni International or distributor.

PTFE bearing wears quickly

- Fluid level may be too low in the tube.

- Immerse the probe deeper into the fluid.

Excessive splashing in sample tube

- Fluid level too low for tube size

Содержание Mixer

Страница 1: ...Omni Mixer Macro User Manual OMNI MIXER HOMOGENIZER www omni inc com...

Страница 2: ...SAVE THIS OWNER S MANUAL The Omni Mixer and Omni Macro has been engineered for economical functionality as well as safety however basic safety precautions and common sense must always be demonstrated...

Страница 3: ...Instruction Manual 17258 17258 1 Overview 1 2 3 4 5 1 Support Post 2 Coupling Adapter 3 Support Base 4 Clamping Knob Assembly 5 Motor Support Housing Description P N QTY Gen Probe Multi Tool 00 402 2...

Страница 4: ...robes can be used in open containers or in sealed chamber assemblies Sample process ing efficiency is effected by Amount of material processed vs size and speed of the generator probe Container geomet...

Страница 5: ...ot thread fully into the generator probe adapter then unscrew the generator probe from the generator probe adapter Slightly rotate the rotor shaft collar on the generator probe Then return to step 2 4...

Страница 6: ...eleased Determining Motor Speed To determine motor speed during operation a tachometer not in cluded may be inserted into the Omni Mixer or Omni Macro through the top of the motor as shown by removing...

Страница 7: ...g wear or part failure NOTE Inspect glass or plastic chambers for any scratches or cracks and do not use if any are found Eye protection and utilization of a secondary container is recommended to prev...

Страница 8: ...n the assembly securely 2 Attach the chamber assembly to the motor by screw ing the chamber cover into the coupling adapter 3 If the chamber assembly does not thread fully into the coupling adapter th...

Страница 9: ...commutator and should be re placed check both brushes since both brushes generally do not wear at the same rate If one brush is worn replace both brushes Make sure that the brushes are installed prop...

Страница 10: ...d turned on but is not functioning Check power cord connectors Check wall socket for power to the outlet Motor is turned ON and makes a buzz ing sound but is not working Brushes may need to be replace...

Страница 11: ...This page left blank intentionally...

Страница 12: ...PN 17258 Rev B 935C Cobb Place Blvd Kennesaw GA 30144 800 776 4431 770 421 0058 www omni inc com...

Отзывы: