MPD 800 User Manual
20
OMICRON
► Do not exceed 7 A
RMS
reactive current at the
IN
input.
For low frequencies below 1 kHz, the input impedance is in the < 100 mΩ range. For high frequencies
above 10 kHz, the termination impedance is equivalent to the termination impedance of 50 Ω connected
to the
PD
output of
CPL1
. Special versions of
CPL1
for NEMA and CISPR measurements feature an
input impedance of 150 Ω or 300 Ω respectively.
Rear panel connections
CPL2
Figure 2-8:
CPL2
rear panel
The GND wing nut is connected to the ground with a short connection.
► Do not ground the
MPD 800
acquisition unit or its accessories in this case.
DANGER
Death or severe injury caused by high voltage or current
A failure inside
CPL1
may result in residual charges at the coupling capacitor or test
object, even with high-voltage side grounded.
► Always observe the five safety rules (see 1.2.2 "Safety rules" on page 6) before
connecting or integrating
CPL1
into a system.
CAUTION
Minor or moderate injury caused by voltage or current possible
If the ratings are violated, damage or loss of precision of the equipment might be the
consequence.
► Observe the maximum current and power ratings of the measurement input.
► Never connect a voltage source directly to the
IN
input.
WARNING
Death or severe injury caused by high voltage or current possible
CPL1
is not intended to be used for measurements which are connected directly to
mains circuits (IEC61010-2-030).
► Make sure to connect
CPL1
to the test setup using appropriate rated accessories
because
CPL1
is not able to handle power frequency fault currents.