Technical data
63
8
Technical data
8.1
Guaranteed values
The values are valid for the period of one year after factory calibration, within 23 °C ±5 °C at
nominal value and after a warm-up time greater than 25 min.
The given input/output accuracy values relate to the range limit value (% of range limit value).
8.2
Power supply
Main power supply
Connection
Connector according to IEC 60320
Voltage, single phase
nominal voltage
operational range
100 ... 240 V AC
85 ... 264 V AC
Power fuse
T 12.5 AH 250 V (5 x 20 mm)
"Schurter", order number 0001.2515
Nominal current
10 A
Frequency
nominal frequency
operational range
50/60 Hz
45 ... 65 Hz
Overvoltage category
II
8.3
Analog measurement inputs
Analog inputs: ANALOG INPUT 1...10
Number of analog measurement inputs 10
Measurement ranges
(RMS value of the sinusoidal shaped
input signal)
10 mV, 100 mV, 1 V, 10 V, 100 V, 600 V
Crest factor
1.75
Sampling frequency
10 kHz
40 kHz
Input impedance
(1 MΩ ±2%) || (170 pF ±50 pF)
Measurement category
CAT II / 600 V
CAT III / 300 V
CAT IV / 150 V
Temperature drift
<25 ppm/K
Содержание DANEO 400
Страница 1: ...DANEO 400 User Manual...
Страница 5: ...Contents 5 Glossary 75 Support 77 Index 79...
Страница 8: ......
Страница 60: ......
Страница 74: ......
Страница 76: ......
Страница 78: ......
Страница 80: ...DANEO 400 80 Time 59 TransView 53 W Web interface 57 wiring the terminals 9...