background image

 
 

 

 

WARNINGS 

Failure to read these warnings may cause the unit to fail, personal injury or property 
damage. 

  Connect the unit to a grounded shockproof socket only. Ensure that the values indicated 

on the nameplate of the unit must correspond with the available connecting conditions. 
The unit must be connected to a single phase, 120 volt AC grounded electrical outlet.  

  Check the ultrasonic unit for possible transport damages before initial operation.  

  In case of visible damage 

do not

 connect the unit to a power source, contact your 

supplier.  

 

Do not use this machine for cleaning live animals, pets or plants.

 

 

Do not plug equipment into a power source that utilizes a GFI receptacle. Ultrasound 
passes a small, trickle current through the neutral which will cause GFI’s to trip. 

 

 

Never plug in or operate the unit (heat or ultrasound) without the appropriate liquid 
level in the tank (1” from the top).  

  This  ultrasonic  cleaning  unit  has  been  designed  for  cleaning  items  using  liquid  soap 

only.  

 

Any detergents or chemicals used in this equipment must be compatible with 300 series 
stainless steel. Do not use any chemicals that contain any strong acids i.e. hydrochloric, 
sulfuric or muriatic acid. These chemicals will cause permanent damage to the stainless 
steel welds.    

 

Use only biodegradable cleaning agents.  Never use solvents or flammable cleaning 
solvents without approval from Omegasonics. Any chemistry with a flash point below 
82˚C should never be used with an ultrasonic cleaner.  

 

  Ensure that sensitive surfaces are exposed to ultrasonic activity for short periods of time 

only.  If  in  doubt  check  the  cleaning  progress  regularly  and  observe  the  state  of  the 
surface material. 

 

Due to the heated liquid in the tank, use the basket to insert or remove parts from the 
tank. Always switch off the unit when placing items in or taking them out of the tank.

 

 

Do not operate the unit with wet hands. 

 

 

Do not open the internal circuitry of the equipment, disassemble any part or parts, or 
move or remove any components or electrical devices. 

 

 

Never attempt to perform maintenance on the equipment when the unit is energized or 
when the cleaning solution is hot. The cleaning liquid and unit heat-up even if heat is 
switched off.  

 

 

Disconnect the power source when moving the unit to a new location.

 

 

Avoid splashing water outside the tank.

 

 
 

 

The manufacturer cannot be held liable for damages of persons, equipment or items cleaned caused 
by improper use

Only qualified technically trained personnel should perform any electrical 

maintenance on this machine.

 

 

 

Содержание 7850TT

Страница 1: ...1 Omegasonics Ultrasonic Cleaning Equipment Operation Instruction Manual OMEGASONICS Table Top Unit 2017 and After Model 7850TT 7950TT Read all instructions thoroughly before operating this equipment...

Страница 2: ...Getting to Know Your Unit 5 Cleaning Solution Limitations 9 Set Up 10 Cleaning 10 Specific Cleaning Procedures 11 Equipment Maintenance 12 Service Life of Transducer Tank 14 Trouble Shooting 15 Omega...

Страница 3: ...d cleaning speed and consistency when compared with simple soaking or immersion with agitation Additionally the bubbles are small enough to penetrate even microscopic crevices cleaning them thoroughly...

Страница 4: ...c or muriatic acid These chemicals will cause permanent damage to the stainless steel welds Use only biodegradable cleaning agents Never use solvents or flammable cleaning solvents without approval fr...

Страница 5: ...een completed Use the permanent ON position for continued operation This requires that the unit be switched off by hand For safety reasons the unit is automatically switched off after 8 hours of perma...

Страница 6: ...heating is switched on Button Turns the ultrasound on off Ultrasound LED Display is lit when ultrasound is switched on To set the value of either the time or heat turn the knob clockwise To reset the...

Страница 7: ...hese can destroy the stainless steel tank by crevice corrosion within a very short period of ultrasonic operation Other media which can destroy the stainless steel tanks when used in high concentratio...

Страница 8: ...f the lid Do not exceed a temperature of 80 C inside the tank CLEANING Timed Select the required temperature Set the required ultrasonic cleaning period When the set cleaning period has run down the u...

Страница 9: ...ean with 64 parts clean water to a temperature of 46 C A higher dilution may be required depending on the severity of contamination If it exists excess water should be removed as quickly as possible U...

Страница 10: ...OmegaSupreme with 10 parts clean water to a temperature of 66 C Place the metal parts into the basket with the cavity containing heaviest residue facing the bottom of the basket Submerge the basket i...

Страница 11: ...t is 46 49 C Blow off excess moisture with compressed air and dry using a hand held hot air dryer Lubricate using light gun oil or silicone spray Test using industry standards Bicycle Parts In the ult...

Страница 12: ...electrical cable for damage regularly in order to prevent electrical accidents Draining Tank Turn equipment off and unplug the power cord Wait at least twenty 20 minutes after the heat is turned off...

Страница 13: ...lation inside the unit Do not put the unit in or under water If the unit is used for medical and sanitary purposes it is necessary to disinfect the transducer tank and the surfaces regularly using sta...

Страница 14: ...even more we recommend observing the following instructions Regularly remove any cleaning residues in particular metal particles and rusty film Use suitable cleaning chemicals with particular caution...

Страница 15: ...h on the unit at on off key Key ultrasound not pressed Switch key on Faulty electronics Return unit to supplier or manufacturer No ultrasound LED display ultrasound is blinking rapidly fault indicatio...

Страница 16: ...revolution of cleaning liquid e g switch on ultrasound Unit produces boiling noise during heating up No revolution of cleaning liquid e g switch on ultrasound Set temperature is exceeded Temperature...

Страница 17: ...osphate rust inhibitors or with another cleaning agent containing rust inhibitors OmegaCitriSurf 2250 Designed for industrial and aerospace applications where stainless steel passivation the removing...

Страница 18: ...tion erosion misuse abuse or transportation damage This warranty is limited to the original purchaser and is not transferable Total liability for any reason whatsoever shall not in any case exceed the...

Отзывы: