11 

 

Firearms 

 

In the ultrasonic bath, preheat 1 part OmegaBlue with 32 parts clean water to a 
temperature of 46˚C.  

 

Disassemble the firearm into its major components. If it is equipped with a scope, it can 
be left attached.  

 

Spray the components with a spray bottle containing 1 part OmegaBlue, 21 parts clean 
water solution. Let sit for 1 – 3 minutes.  

 

Rinse off the residue loosened by the solution.  

 

If the stock is synthetic, it can be cleaned with the other metal parts. If it is wooden, use 
a basket to submerge it into the bath and clean for 1 -2 minutes.  

 

Using a basket, submerge the components into the bath and clean for 1 – 5 minutes, 
until all residues are removed.   

 

You may spot clean using the spray bottle solution and a 3M pad that is green or light in 
color. 

 

Rinse all components with de-ionized water that is 46 - 49˚C. 

 

Blow off excess moisture with compressed air and dry using a hand-held hot air dryer.  

 

Lubricate using light gun oil or silicone spray. 

 

Test using industry standards.  

 
 
 

Bicycle Parts 

 

In the ultrasonic bath, preheat 1 part OmegaSupreme with 10 parts clean water to a 
temperature of 60˚C for aluminum, or 66˚C for chains, derailleurs and gears.  

 

Place the items into the basket with the cavity containing heaviest residue facing the 
bottom of the basket.  

 

Clean the part(s) for 5 minutes. 

 

Once completed, rotate the part 180˚ if the part facing opposite the ultrasonic 
transducers is heavily soiled.  

 

If necessary, clean the part for an additional 5 minutes.  

 

Once the part is cleaned, remove from the bath and rinse using hot water. 

 

Blow off the excess water using compressed air.  

 

Due to the heat of the metal, after being removed from the bath, it will dry quickly, on 
its own (“flash dry”).   

 

Note: If your aluminum parts develop any type of scarring or starbursts on the surface, 
lower the bath temperature to 54˚C. 

Содержание 7850TT

Страница 1: ...1 Omegasonics Ultrasonic Cleaning Equipment Operation Instruction Manual OMEGASONICS Table Top Unit 2017 and After Model 7850TT 7950TT Read all instructions thoroughly before operating this equipment...

Страница 2: ...Getting to Know Your Unit 5 Cleaning Solution Limitations 9 Set Up 10 Cleaning 10 Specific Cleaning Procedures 11 Equipment Maintenance 12 Service Life of Transducer Tank 14 Trouble Shooting 15 Omega...

Страница 3: ...d cleaning speed and consistency when compared with simple soaking or immersion with agitation Additionally the bubbles are small enough to penetrate even microscopic crevices cleaning them thoroughly...

Страница 4: ...c or muriatic acid These chemicals will cause permanent damage to the stainless steel welds Use only biodegradable cleaning agents Never use solvents or flammable cleaning solvents without approval fr...

Страница 5: ...een completed Use the permanent ON position for continued operation This requires that the unit be switched off by hand For safety reasons the unit is automatically switched off after 8 hours of perma...

Страница 6: ...heating is switched on Button Turns the ultrasound on off Ultrasound LED Display is lit when ultrasound is switched on To set the value of either the time or heat turn the knob clockwise To reset the...

Страница 7: ...hese can destroy the stainless steel tank by crevice corrosion within a very short period of ultrasonic operation Other media which can destroy the stainless steel tanks when used in high concentratio...

Страница 8: ...f the lid Do not exceed a temperature of 80 C inside the tank CLEANING Timed Select the required temperature Set the required ultrasonic cleaning period When the set cleaning period has run down the u...

Страница 9: ...ean with 64 parts clean water to a temperature of 46 C A higher dilution may be required depending on the severity of contamination If it exists excess water should be removed as quickly as possible U...

Страница 10: ...OmegaSupreme with 10 parts clean water to a temperature of 66 C Place the metal parts into the basket with the cavity containing heaviest residue facing the bottom of the basket Submerge the basket i...

Страница 11: ...t is 46 49 C Blow off excess moisture with compressed air and dry using a hand held hot air dryer Lubricate using light gun oil or silicone spray Test using industry standards Bicycle Parts In the ult...

Страница 12: ...electrical cable for damage regularly in order to prevent electrical accidents Draining Tank Turn equipment off and unplug the power cord Wait at least twenty 20 minutes after the heat is turned off...

Страница 13: ...lation inside the unit Do not put the unit in or under water If the unit is used for medical and sanitary purposes it is necessary to disinfect the transducer tank and the surfaces regularly using sta...

Страница 14: ...even more we recommend observing the following instructions Regularly remove any cleaning residues in particular metal particles and rusty film Use suitable cleaning chemicals with particular caution...

Страница 15: ...h on the unit at on off key Key ultrasound not pressed Switch key on Faulty electronics Return unit to supplier or manufacturer No ultrasound LED display ultrasound is blinking rapidly fault indicatio...

Страница 16: ...revolution of cleaning liquid e g switch on ultrasound Unit produces boiling noise during heating up No revolution of cleaning liquid e g switch on ultrasound Set temperature is exceeded Temperature...

Страница 17: ...osphate rust inhibitors or with another cleaning agent containing rust inhibitors OmegaCitriSurf 2250 Designed for industrial and aerospace applications where stainless steel passivation the removing...

Страница 18: ...tion erosion misuse abuse or transportation damage This warranty is limited to the original purchaser and is not transferable Total liability for any reason whatsoever shall not in any case exceed the...

Отзывы: