Find below instructions on how to perform a Turbidity Reading. Before starting, verify if the bubble
trapper is clean, if the drain hose is connected and “open” and if the sample is flowing correctly.
Finally insert the sensor unit at the bubble trapper and lock it by turning it counter clockwise
(45degrres).
TURBIDIMETER
/SET UP
READ
TURBIDIMETER
READ
/CALIBRATE
GO TO SAMPLE!
READY?
================
-> 0.0035 NTU
STAND BY
================
-> 8.5100 NTU
S1:» S2:»
-> 4.83 mA
ON LINE CALIBR.
8.5100 NTU
<
>
================
-> 8.5160 NTU
12.
Equipment Operation - Read
22
Press
until
flashes then press
key.
<SEL>
Read
<ENT>
Press
until
flashes then press
key.
<SEL>
Read
<ENT>
Verify if sample is flowing and drain hose is connected,
then press
<ENT>
key to start Reading Mode.
S1:» S2:»
-> 0.0035 NTU
Indicates when the
Bargraph was not chosen
during Set Up Mode.
Press
key to place the equipment in Stand By.
<ENT>
Press
key to exit Stand By and return to Read Mode.
<ENT>
Press
and it will be able to adjust the value
Read, if chosen this procedure during Set Up Mode
(On Line Calibration-page 17). Refer to page 16
(Note A) for instructions on how to adjust this value.
<SEL>
Press
key to verify the Contact and Current
conditions.
<SEL>
In order to exist the reading Mode, press and hold
<ESC>
key. User will then access the main menu.
TURBIDIMETER
/SET UP
READ