DISPLAY CONFIG.?
/
YES
NO
BARGRAPH?
/
YES
NO
Min:0.0000NTU
Max:10.0000NTU
<
<
>
>
CONTACTS CONFIG?
/
YES
NO
CONFIG. CONTACT
S1
CONFIG. S1:
/OFF
ON
SET POINT S1:
5.0000 NTU
<
>
ACTING:
/REVERSE
DIRECT
HYSTERESE:
0.0000NTU
<
>
BURN-OUT CONFIG.
STAND-BY:
HOLD
CONFIG. CONTACT
S2
CONFIG. S2:
/OFF
ON
From Page 17
A
18
12.
Equipment Operation - Set Up (cont.)
User can program the display to show information like
Barr graph, Clock and more.
User can choose if desire to have Barr graph shown
above the Reading screen. When Bar Graph is
displayed, the information about Sensibility and Sample
Temperature will not be displayed!
User can now adjust the Minimum and Maximum
values for the Bar Graph. Refer to Page 16 (Note A) for
instructions on how to modify these values.
User has the option to configure Contacts SP 1and SP 2.
Press
key to choose the desired option then
press
key to confirm.
<
SELECT>
<ENTER>
If user chooses Contact S1 as On, while at Reading Mode,
a or a will be displayed after the S1, indicating that
this Contact is On. If user chooses Contact S1 as Off, while
at Reading Mode, a will be
X
User can adjust the value for SP 1.
Refer to Page 16 (Note A) for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
User can choose the Acting desired.
Press
key to choose the desired option then
press
key to confirm
<
SELECT>
<ENTER>
User can adjust the value for Hystereses.
Refer to Page 16 (note A)for instruction on how to adjust
this value. Resolution will depend upon the Resolution
chosen while at Set Up.
For the Burn Out configuration, user will have three options as:
On - Contact is going to stay On all the time, Off - contact will stay
Off all the time or Hold - contact is going to follow the last situation,
before going to Hold status.
Press
key until the desired option flashes, then press
to confirm.
<SEL>
<ENT>
If user chooses Contact S2 as On, while at Reading Mode,
a or a will be displayed after the S1, indicating that
this Contact is On. If user chooses Contact S1 as Off, while
at Reading Mode, a will be
X