Commissioning and operation
Series TCL-3M165E2
- 22 -
Drain syringe and magnetic lifter:
The drain syringe is used for pumping out the calibration liquid from the tank of the micro
bath. The magnetic stirrer is removed with the help of the magnetic lifter.
Both the activities are required before another adapter sleeve or calibration insert is inserted
in the calibrator.
Tub insert (optional / accessories):
We recommend using a tub insert if you
•
frequently change between dry block, infrared, surface and micro
bath function.
•
frequently work with different calibration liquids.
The tub insert is placed into the tank with the help of the sleeve exchange
tool.
Just like the tank, the tub insert can be closed with the associated cover.
Both threaded covers are leakproof, so the calibration liquid can be left in
the tank or the tub insert during transport.
4.3.4.1 Notes on calibration liquid
Different calibration liquids supply varying calibration results due to their specific
characteristics. Adjustment to the respective calibration liquid has to be carried out by the
manufacturer.
In order to achieve the best possible accuracy of a micro bath, it has to be filled with a
suitable calibration liquid.
The calibration liquid is poured directly into the tank or a tub insert.
When using water as the calibration liquid:
Only use distilled water, otherwise excessive limescale and soiling will build up in the
tank.
When using silicone oil as the calibration liquid:
Wear safety goggles for the eyes without fail when working with silicone oil!
Use only the silicone oil recommended here.
Always read the safety data sheet supplied with the silicone oil before using it.
Always ensure adequate ventilation when working with silicone oil, since hazardous
substances can be released.
Spilled or leaked silicone oil results in an extreme danger of slipping. Clean the
affected ranges by suitable means.
Since silicone oil is hygroscopic, always use the associated transport cover to
close the tank or the tub insert after use.
Содержание TCL-3M165E2 Series
Страница 1: ...Series TCL 3M165E2 Temperature Calibrator...
Страница 2: ...Series TCL 3M165E2 2...
Страница 58: ...For your notes 58 For your notes...
Страница 59: ...Series TCL 3M165E2 Technical changes reserved 59...
Страница 60: ...Series TCL 3M165E2 60 M 5663 0318...