Series TCL-3M165E2
Operation
Technical changes reserved
- 39 -
Change test point:
Tap on the
icon to modify the selected test point.
The window for the test point setting is opened.
Change the temperature and the dwell time and confirm the new values.
The new values are adopted and you will be returned to the previous window.
Test point calculation
Here, you can easily have the steps between two temperatures calculated. If a change is
made to a parameter, the other parameters dependent on it are calculated afresh
automatically.
Tap the parameter field to open the window "Calculate test point".
Step, one-sided
:
Activate the parameter field by tapping if the test points are to be approached in
succession from the first to the last test point.
Step, two-sided
:
Activate this parameter field by tapping if the test points are to be approached in
succession from the first to the last back to the first test point.
Define first and last test point
To do so, tap on the parameter fields, input the new values and confirm the input.
The new values are adopted, the interval is calculated afresh and you will be returned to
the previous window.
Cycles:
You can set the number of cycles here. For every cycle, the test points in the selection list
are approached successively.
Tap on the parameter field to change the number of cycles.
Input the desired number in the input window and confirm the input.
The new values are adopted and you will be returned to the previous window.
Set interval:
For fixed intervals, tap on the parameter field "Interval", input the desired step width and
confirm the input.
The new interval is adopted and the number of steps is calculated. If necessary, the last
test point is also adjusted to match, and you will be returned to the previous window.
Defining the number of steps:
Note in this context that the number of test points is greater than the number of steps by 1.
Tap on the parameter field "Number of Steps" to change the number of steps.
Input the desired number and confirm the input.
The number of steps is adopted and the interval is calculated afresh.
Set dwell time:
Tap on the parameter field "Dwell time", input the new values and confirm the input.
The new value is adopted and you will be returned to the previous window.