INST.No.INE-418-1P1CE
-25-
Phenomenon Content
Countermeasure
6)
“Er4”
is displayed in the
main display. 2
The batteries are dead
.
If
“Er4”
is still displayed after the
battery replacement, this meter has
been abnormal.
Replace the batteries with new ones. (Ref. [5.1 Loading of batteries])
If
“Er4”
is still displayed after the battery replacement, contact your
OMEGA’s agent
.
7) Nothing including the clock
is displayed.
No power has been supplied.
Replace the batteries with new ones.
(Ref. [5.1 Loading of batteries])
8) None displayed
The batteries are dead
Replace the batteries with new ones.
(Ref. [5.1 Loading of batteries])
9)
“oFL”
is displayed in the
main
display 1.
Temperature measured value
overflowed
Check the temperature measuring range.
Check the connection of the temperature/humidity sensor.
(Ref. [7. Replacement of temperature/humidity sensor])
10)
“uFL”
is displayed in the
main display 1.
Temperature measured value under
flowed
Check the temperature measuring range.
Check the connection of the temperature/humidity sensor.
(Ref. [7. Replacement of temperature/humidity sensor])