3. When

 

Meter shows erratic readings, it must be sensor failed or power is weak. 

4. Only select one of the two electrodes while measuring the same water zone, otherwise meter appears erratic 

readings. Read two parameters at the same time are only available for measuring two different water sources.

 

(1) POWER 
NOTE: Make sure you have connected the electrode to the meter before power on. 

Momentarily press 

PWR

 button to turn on the meter, press and hold 

PWR

 button to turn off meter. 

 
(2) Check PH mV value 

Short press

 MODE

 to check pH in mV unit. 

 

(3) CHANGE THE TEMP. UNIT TO 

/

 

Short press 

UNIT

 button to toggle 

  or 

 

(4) HOLD 

Freeze the current readings of pH

, then icon “

Hold

” will appear on the top left of screen. 

 

(5) AUTO POWER OFF 

Meter will turn off automatically in 15 minutes when no use, to disable auto power off function by pressing

 

HOLD 

and

 PWR

 

button, “n” momentarily shows on screen, now meter is at non-sleep mode, then turns to 

normal measurement, Meter default auto power off. 
 

(6) FACTORY RESET 

a) 

pH electrode 

Whenever replacing a new pH electrode. Long pressing 

MODE+CAL

 to enter pH calibration mode then strongly 

recommend doing factory reset by pressing 

SET+UNIT 

under pH4/7/10 calibration mode, screen will   

show “

rFS

” momentarily. 

 

(7) TEMPERATURE PROBE TYPE SELECTION 
NOTE: 

If using NTC 10K probe but select 30K type. The temperature value will not be accurate. 

Default setting is NTC 10K. User can change it depends on the electrode type to get the correct value. 

NTC 10K: 

Negative Temperature Coefficient 25

  = 10 K 

NTC 30K: 

Negative Temperature Coefficient 25

  = 30 K 

NOT: 

External temperature probe is excluded, user can enter the temperature degree by their own 
temperature instrument, default: 25

    Adjustable range: 0.0

~90.0

 

Step 1: 

Must

 

select correct probe type before measurement, otherwise the value would be incorrect.

 

Step 2:

 Long pressing 

UNIT

 button, the meter default is 

ntc 10k

, short pressing 

UNIT 

button to toggle ntc 30k   

  not. 

Step 3: 

Long pressing 

UNIT 

button again to save the setting, meter shows 

“SA”

 at the bottom of LCD and then   

return to normal measurement mode. 

 

 

 

 

 

 

 

 

 

Содержание PHH444

Страница 1: ...e mail info omega com For latest product manuals omega com en us pdf manuals Shop online at omega com User sGuide PHH444 Handheld pH mV Meter and pH Electrode Kit TM 2...

Страница 2: ...pecifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Servi...

Страница 3: ...brated before shipment Warranty period of meter is two years Electrode is one year SUPPLIED Meter Electrode x 1 pc pH Battery AAA x 3 pcs pH buffer solution x 3 pcs pH 4 pH7 pH10 Soaking Solution x 1...

Страница 4: ...rt the braid into the center of the connector Continuing pushing the braid into the connector until it will not go in any farther Do this gently and slowly do not bend the braid c Turn the BNC connect...

Страница 5: ...measurement Meter default auto power off 6 FACTORY RESET a pH electrode Whenever replacing a new pH electrode Long pressing MODE CAL to enter pH calibration mode then strongly recommend doing factory...

Страница 6: ...ibrating the pH in the incorrect buffer solution Er 3 will appear on the bottom of screen 4 User can do 1 2 or 3 points calibration Rinse the probe with de ionized water before doing each pH buffer ca...

Страница 7: ...NIT button to select not type HOLD to increase ADJ to decrease SET to left digit CAL to right digit Long pressing UNIT again to save setting until screen shows SA and back to normal measurement mode 4...

Страница 8: ...umulate when water in a salt solution evaporates during contact with air These salt crystals can be removed by rinsing the area with warm water TROUBLE SHOOTING Q1 pH display erratic readings or calib...

Страница 9: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Отзывы: