KEY PAD (CONTROLS) 

PWR: 

Power on (Press in one second) or power off (Press more than 2 seconds when operation). 

SET: 

Move to left digit. (Under setting mode). Long pressing to save setting or calibrate the reading. 

CAL: 

Move to right digit. (Under setting mode). 

MODE: 

Check pH in mV measurement. Select pH4, pH7, pH10 calibration mode. 

UNIT: 

Short pressing to change temperature unit 

/

. Long pressing to select temperature electrode type. 

Short pressing to select NTC: Negative Temperature Coefficient)/ NOT: no remote temperature probe. 
Long pressing again to save electrode type. 

HOLD: 

Freeze current readings (

Hold

 shows at top of LCD). Increase value. (Under setting mode). 

ADJ: 

Decrease value. (Under setting mode). 

SET+UNIT: 

Recover to factory setting. (Under pH calibration mode) 

MODE+CAL: 

pH Calibration. 

HOLD+PWR: 

Disable Auto Power Off. 

OPERATION 

pH INSTALLATION (BNC connector): 

a)

Insert the pH electrode into the top of right hole. And insert the 3.5mm Ø  diameter phone jack ATC sensor

parts into the middle of hole.

b)

Hold the BNC connector in one hand; with the other, insert the braid into the center of the connector.

Continuing pushing the braid into the connector until it will not go in any farther. Do this gently and slowly;

do not bend the braid.

c)

Turn the BNC connector in a clockwise direction, until you cannot turn it any more.

NOTE: Remove batteries when not use! 

Power-off switch:

 When the meter is turned off, the internal CPU does not shut down completely, it will keep 

detecting the buttons per milliseconds, letting meter know if the user wants to activate the meter or not. It will 

consume the power by each detecting, in order to save the power, you may pull down the switch.   

NOTE: 
For each operation, make sure you are using fresh batteries, same brand, same power of batteries is 

required, and 

otherwise meter shows erratic readings or leakage. Warranty is void if not follow the notices. (Note

remove batteries 

when not use

1. Power supply: 
(a) AAA batteries x 3pcs. 
(b) Battery icon 

indicates when power is weak, replace with new batteries immediately as the readings 

now on 

LCD are incorrect due to weak power. 

Battery life: Approx. 50 hours for continuous used.

2. Make sure sensor electrode 

and meter are well connected. Don’t attempt to detach sensor probe from meter

while in operation.

Pull down 
the switch 

Содержание PHH444

Страница 1: ...e mail info omega com For latest product manuals omega com en us pdf manuals Shop online at omega com User sGuide PHH444 Handheld pH mV Meter and pH Electrode Kit TM 2...

Страница 2: ...pecifications without notice Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Servi...

Страница 3: ...brated before shipment Warranty period of meter is two years Electrode is one year SUPPLIED Meter Electrode x 1 pc pH Battery AAA x 3 pcs pH buffer solution x 3 pcs pH 4 pH7 pH10 Soaking Solution x 1...

Страница 4: ...rt the braid into the center of the connector Continuing pushing the braid into the connector until it will not go in any farther Do this gently and slowly do not bend the braid c Turn the BNC connect...

Страница 5: ...measurement Meter default auto power off 6 FACTORY RESET a pH electrode Whenever replacing a new pH electrode Long pressing MODE CAL to enter pH calibration mode then strongly recommend doing factory...

Страница 6: ...ibrating the pH in the incorrect buffer solution Er 3 will appear on the bottom of screen 4 User can do 1 2 or 3 points calibration Rinse the probe with de ionized water before doing each pH buffer ca...

Страница 7: ...NIT button to select not type HOLD to increase ADJ to decrease SET to left digit CAL to right digit Long pressing UNIT again to save setting until screen shows SA and back to normal measurement mode 4...

Страница 8: ...umulate when water in a salt solution evaporates during contact with air These salt crystals can be removed by rinsing the area with warm water TROUBLE SHOOTING Q1 pH display erratic readings or calib...

Страница 9: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Отзывы: