background image

HHTQ35 Digital Torque Gauges 

 

User’s Guide 

 
 

 

 

 

8

4.3 Menu navigation basics 

Most of the gauge’s various functions and parameters are configured through the main menu. To access 
the menu press 

MENU

. Use the 

UP

 and 

DOWN

 keys to scroll through the items. The current selection is 

denoted with clear text over a dark background. Press 

ENTER

 to select a menu item, then use 

UP

 and 

DOWN

 again to scroll through the sub-menus. Press 

ENTER

 again to select the sub-menu item.  

 
For parameters that may be either selected or deselected, press 

ENTER

 to toggle between selecting and 

deselecting. An asterisk (

*

) to the left of the parameter label is used to indicate when the parameter has 

been selected. 
 
For parameters requiring the input of a numerical value, use the 

UP

 and 

DOWN

 keys to increment or 

decrement the value. Press and hold either key to auto-increment at a gradually increasing rate. When 
the desired value has been reached, press 

ENTER

 to save the change and revert back to the sub-menu 

item, or press 

ESCAPE

 to revert back to the sub-menu item without saving. Press 

ESCAPE

 to revert one 

step back in the menu hierarchy until back into normal operating mode. 
 
Refer to the following sections for details about setting up particular functions and parameters. 
 
 

5   OPERATING MODES 

 

Caution! 

In any operating mode, if the capacity of the instrument has been exceeded by more than 110%, 
the display will show “OVER” to indicate an overload. A continuous audible tone will be sounded 
(if beeps are enabled) until the MENU key has been pressed or the load has been reduced to a 
safe level. 

 
Three operating modes are possible with the HHTQ35 torque gauge. To cycle between the modes, press 

MODE

 while in the home screen. 

 
5.1 Real time (RT) 

The primary reading corresponds to the live measured reading.  
 

5.2 Peak Clockwise (PCW)  

The primary reading corresponds to the peak clockwise reading observed. If the actual load decreases 
from the peak value, the peak will still be retained in the primary reading area of the display. Pressing 

ZERO

 will reset the value. 

 

5.3 Peak Counter-clockwise (PCCW)

  

Same as above, but for counter-clockwise readings. 
 
 

Содержание HHTQ35

Страница 1: ...HHTQ35 Digital Torque Gauges...

Страница 2: ...HHTQ35 Digital Torque Gauges User s Guide 1...

Страница 3: ...operation instructions Dimensions and specifications are also provided For additional information or answers to your questions please do not hesitate to contact us Our technical support and engineeri...

Страница 4: ...ng operation 1 Never operate the gauge if there is any visible damage to the AC adapter or the gauge itself 2 Ensure that the gauge is kept away from water or any other electrically conductive liquids...

Страница 5: ...s 1 When battery life is greater than 75 the following indicator is present 2 When battery life is between 50 and 75 the following indicator is present 3 When battery life is between 25 and 50 the fol...

Страница 6: ...nt Side loading or off center loading can damage the instrument whether power is turned on or off 3 2 Mounting to a plate The indicator portion of the gauge can be mounted to a plate with four thumb s...

Страница 7: ...left indicates counter clockwise load If set points are enabled triangular markers are displayed for visual convenience This indicator reflects the actual load which may not correspond to the primary...

Страница 8: ...el Secondary Function Powers the gauge on and off Press briefly to power on press and hold to power off Active only when the home screen is displayed ENTER Various uses as described in the following s...

Страница 9: ...ert back to the sub menu item or press ESCAPE to revert back to the sub menu item without saving Press ESCAPE to revert one step back in the menu hierarchy until back into normal operating mode Refer...

Страница 10: ...e average of a single value is the value itself To access digital filter settings select Filters from the menu The display will appear as follows Two filters are available Current Reading Applies to t...

Страница 11: ...the current live load 9 COMMUNICATIONS Communication with the HHTQ35 is achieved through the micro USB port located along the left side of the housing as shown in the illustration in the Power sectio...

Страница 12: ...e connecting an Omega USB device 5 After Windows as restarted plug in the device The following will occur Windows 7 Operating Systems When the Omega USB device has been plugged into a USB port the dri...

Страница 13: ...Torque Gauges User s Guide 12 Select No not this time then click Next 6 The next screen appears as follows Select Install the software automatically Recommended then click Next 7 The next screen appe...

Страница 14: ...d 9 2 Communication Settings To set up communication settings select USB Settings from the menu The screen appears as follows Communication settings are permanently set to the following Data Bits 8 St...

Страница 15: ...ity of the instrument Vertical orientation is preferable to avoid side loading which can affect the readings Suitable certified calibration equipment is required and caution should be taken while hand...

Страница 16: ...ar as follows If failed 5 The following screen appears after the offsets have been calculated Keep the sensor in a vertical position as explained in Section 10 1 Attach weight fixtures brackets hooks...

Страница 17: ...he UP and DOWN keys to adjust the torque value as required The torque values default to even increments as indicated by the previously entered number of data points described in Step 1 Then press ENTE...

Страница 18: ...isplayed at the start of calibration if a disallowed unit is selected Ensure that the load is not swinging oscillating or vibrating in any manner Then try again The calibration load does not match the...

Страница 19: ...e will ignore the Automatic Shutoff setting and remain powered on until the POWER key is pressed 11 2 Backlight Several initial settings are available upon powering on the gauge To access these settin...

Страница 20: ...The Set Point alert can be configured to be either a momentary tone or a continuous tone until the load is restored to a value between the set points To configure the functions for which audible tone...

Страница 21: ...alarms calibration Safe overload 150 of full scale display shows OVER at 110 and above Weight 1 8 lb 0 8 kg Included accessories Carrying case AC adapter battery USB cable resource CD USB driver user...

Страница 22: ...x 0 01 14 x 0 01 HHTQ35 50Z 50 x 0 05 36 x 0 05 36 x 0 05 HHTQ35 12 192 x 0 2 12 x 0 01 135 x 0 1 HHTQ35 50 800 x 1 50 x 0 05 570 x 0 5 HHTQ35 100 1600 x 2 100 x 0 1 1150 x 1 12 4 Dimensions IN MM 3 9...

Страница 23: ...HHTQ35 Digital Torque Gauges User s Guide 22...

Страница 24: ...HHTQ35 Digital Torque Gauges User s Guide 23 M 5086 0712...

Отзывы: