Operating Instructions
3-8
Probe Offset
The probe offset feature compensates for temperature probe errors, significantly improving
overall measurement uncertainty. Probe offset can be set for Channel 1 and 2 individually.
Once set, the probe offset is automatically applied to all subsequent measurements and
statistics on the offset channel.
Current statistics will be invalidated after changing settings such as probe offset. Press
to delete existing statistics data and initiate a new statistics session.
Probe offset rounding errors may occur if temperature units are changed while a
probe offset is active. When using a probe offset, verify and if necessary correct the
programmed probe offset after changing temperature units.
To set the probe offset when using an un-calibrated temperature probe:
1.
Connect the temperature probe to Channel 1 or Channel 2 (as desired) of the
instrument;
2.
Place the probe into a known temperature reference such as a thermowell or ice
bath
3.
Allow the temperature probe to stabilize in the ice bath or thermowell by observing
the instrument trend indicators for the
appropriate channel;
4.
5.
Press
three (3) times to cycle to the
Channel 1 Offset parameter;
6.
Observe the current Channel 1 temperature measurement displayed on the top
measurement line of the display, and current offset value displayed on the second
line of the display;
7.
Press
to set the offset in 0.1 ° increments until the displayed temperature
equals the known temperature reference value;
8.
Press
to save the offset value and proceed to Channel 2 offset (HH932T only),
or press
to save the offset value and exit the Setup Menu.
a.
Alternatively, to disregard the new offset value and exit the Setup Menu,
press
.
9.
OFFSET is displayed at the top-left of the LCD display.
5
Probe offset measurement using an ice bath or thermowell should only be performed by personnel trained
and qualified in the use of such instruments and related metrology methods.
Neither trend indicator is
displayed when the
temperature measurement
has stabilized.
Содержание HH931T
Страница 39: ...Appendices B i B EXPANDED INSTRUMENT UNCERTAINTIES Thermocouple Type B ...
Страница 40: ...Appendices B ii Thermocouple Type E ...
Страница 41: ...Appendices B iii Thermocouple Type J ...
Страница 42: ...Appendices B iv Thermocouple Type K ...
Страница 43: ...Appendices B v Thermocouple Type N ...
Страница 44: ...Appendices B vi Thermocouple Type R ...
Страница 45: ...Appendices B vii Thermocouple Type S ...
Страница 46: ...Appendices B viii Thermocouple Type T ...