13

Warning

Yellow 

LED

Green 

LED

Red 
LED 

Error output

Offset adjustment 
necessary

 1

Flashes

on 

off

off

Alarm

Yellow 

LED

Green 

LED

Red 
LED  

Error output

Interpolator electronics 
errors 

2

Flashes

off

Flashes

Pulse

Error at pick-up 

3

off

on /  off

on 

on

Flow overload 

on

off

on

on

max. frequency range 
exceeded 
(>100 kHz) 

4

on

on

Flashes

Pulse

Fluid temperature 
> 120°C 

5

Flashes

on

Flashes

Pulse

Table 6: Alarm and warning messages

17. Alarm and Warning Messages

The electronics of the FHG flow meters can detect five events that could 
lead to measurement errors. In case of serious errors, the third output has 
a ”high“ signal or a ”pulse“ signal if activated with the bridge B1. The 
different error causes can be determined with the states of the three LEDs.

The red LED is linked with the error output. Each active state of this LED or 
the error output signals an event that has negative effects on the measure-
ments.

Description of the Error Messages

1. Offset adjustment necessary: The sensor and/or the preamplifier elec- 
  tronics were replaced. A different size was set.

2. Electronics errors: Defective component in interpolator circuit, unable to 
  determine internal configuration values

3. Pick-up errors: The sensor is defective or quit working. The distance bet- 
  ween the sensor and the magnet wheel has changed = mechanical  
 damage

4. Overload: The maximum permissible flow range was exceeded

5. Frequency errors: The max. output frequency of 100,000 Hz was excee- 
  ded. The IPF is dimensioned too high for the respective flow

6. Temperature errors: The temperature of the fluid is too high (> 120°C) 
  and may result in flawed or incorrect measurements

18. Preamplifier Technical Data

Scanning sensor

 

2 x GMR sensors in a bridge circuit (sin/cos)

Adjustment

 

automatic offset adjustment

Resolution

 

programmable 1, 2, 5, 10, 25, 32, 50, 64, 100, 128 

Adjustable pulse filtering

 

0Z, 0.25Z, 0.5Z, 0.75Z, 1.0Z, 1.25Z, 1.5Z, 1.75Z, 2.0Z, 2.25Z, 2.5Z, 2.75, 3.0Z, 3.25Z, 3.5Z, 3.75Z (Z: gear unit)

Frequency

 

up to 100 kHz

Output signals

 

Channel A, channel B, directional signal DIREC (high: positive, low: negative) or error signal ERROR 

   

(high or pulse: error)

Channel A and B

 

two signal outputs for outputting the digital flow sensor signals, a channel offset of 90° between channel A and  

   

channel B;

Flow direction

 

Detection of flow direction from the channel offset of the signals from channel A to channel B or via the separate direc- 

   

tional signal.

Outputs

 

Three current-limited and short-circuit-proof power amplifiers (channel A, channel B, DIREC / ERROR); integrated  

   

adjustment to a characteristic impedance of 75 Ω; driver current approx. 300 mA at supply of 24 V; small saturation 

   

voltage of up to 30 mA load current, short switching times, integrated freewheeling diodes against V

b

 and GND, ther- 

   

mal shutdown with hysteresis; in case of error, the outputs are high impedance;

   

The 24 V line drivers are designed for control applications with cable adjustment

Error messages

 

Electronics fault (e.g. faulty interpolator), sensor errors (e.g. sensor failure), offset adjustment necessary, overload (flow  

   

peaks), frequency error (> 100 kHz), temperature error (> 120°C)

Operating voltage

 V

b

 = 10 … 28 VDC

Current consumption

 l

noload

 = approx. 40 mA, total current consumption depends on output load

Содержание FHG 100

Страница 1: ...omega com e mail info omega com For latest product manuals omegamanual info FHG FLOW METERS Shop online at User sGuide...

Страница 2: ...ged by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free 0 800...

Страница 3: ...OPERATING INSTRUCTIONS For FHG Series Flow Meters...

Страница 4: ...ics 11 15 Signaling LEDs 12 16 Operating Mode Messages 12 17 Alarm and Warning Messages 13 18 Preamplifier Technical Data 13 19 Preamplifier Pin Assignment 13 20 Maintenance 14 21 Returning for Repair...

Страница 5: ...ply them by a high resolution interpolator using adjustable settings The square wave signals are bidirectional and can be utilised by any evaluating instrument as well as computers and PLC controls Th...

Страница 6: ...tion check and verify the following properties aspects of the respective circumstances of your system to ensure operation is trouble free safe and reliable 1 The Fluid to be Processed Are the flow met...

Страница 7: ...qualify as pressure equipment as defined by Section 1 Paragraph 2 1 4 of the directive listed above and as such are affected by the regulations of this directive Omega flow meters must therefore meet...

Страница 8: ...Thread Please comply with the screw in depths and sealing systems PTFE tape or liquid sealants such as adhesives are not permitted Fastening The devices must be installed stress free into the pipeline...

Страница 9: ...ply them by a high resolution interpolator using adjustable settings The square wave signals phase shif ted by 90 are bidirectional and can be utilised by any evaluating device as well as computers an...

Страница 10: ...surement volume Vm cm3 Imp K Factor Imp l K Factor Imp gal 1 0 10 00000 100 379 2 1 5 00000 200 757 5 2 2 00000 500 1893 10 3 1 00000 1000 3785 25 4 0 40000 2500 9464 32 5 0 31200 3200 12113 50 6 0 20...

Страница 11: ...9 FHG 1xx2 FLOW vs FREQUENCY FHG 1xx4 FLOW vs FREQUENCY Flow diagrams vs frequency...

Страница 12: ...xample Flow meter FHG 1xx4 Max processable input frequency of the downstream evaluating unit 20 kHz Max operating flow 140 l min Path 1 The diagram yields an IPf of 25 Path 2 IPF fmax x VmIPF1 x 60 Qm...

Страница 13: ...ser is able to set the degree of filtering in the form of partial volumes using rotary coding switches Adjustable pulse filtering 0Z 0 25Z 0 5Z 0 75Z 1 0Z 1 25Z 1 5Z 1 75Z 2 0Z 2 25Z 2 5Z 2 75 3 0Z 3...

Страница 14: ...fly press the S2 key for the acknowledgment The new pulse rate is enabled a once The rotary coding switch for the pulse filtering has 16 switch settings The degree of filtering is determined with quar...

Страница 15: ...tive flow 6 Temperature errors The temperature of the fluid is too high 120 C and may result in flawed or incorrect measurements 18 Preamplifier Technical Data Scanning sensor 2 x GMR sensors in a bri...

Страница 16: ...ly well shielded connection cables with a wire cross section of 4 to 5 x 0 25 mm Please note that the housing of the round pin plug is metallic has a connection for the shield and that the potential o...

Страница 17: ...flow meter has been completely cleaned and flushed This is to protect our employees and makes our work easier In case of noncompliance with this rule the devices are returned to the sender without att...

Страница 18: ...sswiderstand p Durchflusswiderstand p Durchflussbereich 0 bis 50 l min Flow range 0 up to 120 l min Flow range Q Flow range Q Flow range Q Flow range Q Flow range Q Flow range pressure drop p Flow ran...

Страница 19: ...sseinheit AR 2500 R Flanschanschlussma e DIN EN 1092 Gewicht 120 kg FHG1xx5 FHG1xx7 Preamplifier Preamplifier Earthing Earthing Weight 81 kg Weight 81 kg Weight 120 kg Sensor module Sensor module Conn...

Страница 20: ...ier Preamplifier Earthing Earthing on both sides on both sides M8 15 deep M8 15 deep Weight 12 kg Weight 12 7 kg Weight 24 8 kg Weight 22 kg Sensor module Test Port Test Port Sensor module Connection...

Страница 21: ...2 Evaluating unit e g display Flow sensor FHG Q 8l min V 11 53 l direction pos f 424 9 Hz control input 0 voltage blue channel 1 white channel 2 black Error Direction grey 25 Pin Assignment Channel 1...

Страница 22: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Страница 23: ...city Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industria...

Отзывы: