background image

12

During initial startup, the switch S2 must first be set to the corresponding 
preferred direction of the flow. The positive flow direction of the FHG flow 
meter system is specified in the top view of the 5-pin M12 connector. In this 
case, the switch S1 must be set to ON. For the opposite negative direction, 
the switch position is to be down and thus set to OFF. This setting ensures 
that the pulse filtering is activated in the right direction from the very begin-
ning after switching on the supply voltage.

Pin 5 of the M12 connector is either used for the separate direction signal 
or an error signal. This is set accordingly with the bridge B2. The figure 
above depicts the bridge attached to the middle and right pin of the 3-pin 
row of pins, which routes the separate zero signal to the third output. If the 
bridge is on the left and middle pin, the error signal in case of a fault is 
output. A description of the error states is found in the ”Alarm and Warning 
Messages“ chapter. 

Ten different interpolation factors can be set with the coding switch S3. The 
corresponding interpolation factors for the respective switch positions are 
listed in Table 3. This setting can be changed at any time while the system 
is running. Simply use a small screwdriver to adjust the rotary coding switch 
and then briefly press the S2 key for the acknowledgment. The new pulse 
rate is enabled a once.

The rotary coding switch for the pulse filtering has 16 switch settings. The 
degree of filtering is determined with quarter gear division increments. The 
corresponding suppressed partial volumes of the respective size are listed 
in Table 4. Changes can also be performed during operation and become 
active after pressing the S2 key.

The electronics is sensitive to electrostatic discharges. People making  
adjustments to the electronics must first discharge their electrostatic charges 
using a grounded object.

Important:

People making adjustments to the electronics must first discharge their electrostatic 

charges using a grounded object.

15. Signaling LEDs

The signaling LEDs provide information about the corresponding status of the 
electronics. These include certain operating and fault states (see Figure 5). 

The three LEDs have a different combination of states for each signal. The 
LEDs signal either operating modes or alarms and warnings. Operating 
mode messages signal the respective mode that has been set.
Alarm and warning signals provide explicit information about overload, 
conditions that can negatively affect the measurement, or component er-
rors of the measuring system.

 

Fig. 5: Signaling LEDs of the preamplifier electronics

16. Operating Mode Messages

Mode

Yellow 

LED

Green 

LED

Red 
LED 

Error output

Normal operation

off

on

off

off

Offset mode

off

Flashes

off

off

Table 5: Operating mode messages

Green

Yellow

Red

Содержание FHG 100

Страница 1: ...omega com e mail info omega com For latest product manuals omegamanual info FHG FLOW METERS Shop online at User sGuide...

Страница 2: ...ged by the United Kingdom Office Toll Free 0800 466 342 TEL 33 0 161 37 29 00 FAX 33 0 130 57 54 27 e mail sales omega fr Germany Austria Daimlerstrasse 26 D 75392 Deckenpfronn Germany Toll Free 0 800...

Страница 3: ...OPERATING INSTRUCTIONS For FHG Series Flow Meters...

Страница 4: ...ics 11 15 Signaling LEDs 12 16 Operating Mode Messages 12 17 Alarm and Warning Messages 13 18 Preamplifier Technical Data 13 19 Preamplifier Pin Assignment 13 20 Maintenance 14 21 Returning for Repair...

Страница 5: ...ply them by a high resolution interpolator using adjustable settings The square wave signals are bidirectional and can be utilised by any evaluating instrument as well as computers and PLC controls Th...

Страница 6: ...tion check and verify the following properties aspects of the respective circumstances of your system to ensure operation is trouble free safe and reliable 1 The Fluid to be Processed Are the flow met...

Страница 7: ...qualify as pressure equipment as defined by Section 1 Paragraph 2 1 4 of the directive listed above and as such are affected by the regulations of this directive Omega flow meters must therefore meet...

Страница 8: ...Thread Please comply with the screw in depths and sealing systems PTFE tape or liquid sealants such as adhesives are not permitted Fastening The devices must be installed stress free into the pipeline...

Страница 9: ...ply them by a high resolution interpolator using adjustable settings The square wave signals phase shif ted by 90 are bidirectional and can be utilised by any evaluating device as well as computers an...

Страница 10: ...surement volume Vm cm3 Imp K Factor Imp l K Factor Imp gal 1 0 10 00000 100 379 2 1 5 00000 200 757 5 2 2 00000 500 1893 10 3 1 00000 1000 3785 25 4 0 40000 2500 9464 32 5 0 31200 3200 12113 50 6 0 20...

Страница 11: ...9 FHG 1xx2 FLOW vs FREQUENCY FHG 1xx4 FLOW vs FREQUENCY Flow diagrams vs frequency...

Страница 12: ...xample Flow meter FHG 1xx4 Max processable input frequency of the downstream evaluating unit 20 kHz Max operating flow 140 l min Path 1 The diagram yields an IPf of 25 Path 2 IPF fmax x VmIPF1 x 60 Qm...

Страница 13: ...ser is able to set the degree of filtering in the form of partial volumes using rotary coding switches Adjustable pulse filtering 0Z 0 25Z 0 5Z 0 75Z 1 0Z 1 25Z 1 5Z 1 75Z 2 0Z 2 25Z 2 5Z 2 75 3 0Z 3...

Страница 14: ...fly press the S2 key for the acknowledgment The new pulse rate is enabled a once The rotary coding switch for the pulse filtering has 16 switch settings The degree of filtering is determined with quar...

Страница 15: ...tive flow 6 Temperature errors The temperature of the fluid is too high 120 C and may result in flawed or incorrect measurements 18 Preamplifier Technical Data Scanning sensor 2 x GMR sensors in a bri...

Страница 16: ...ly well shielded connection cables with a wire cross section of 4 to 5 x 0 25 mm Please note that the housing of the round pin plug is metallic has a connection for the shield and that the potential o...

Страница 17: ...flow meter has been completely cleaned and flushed This is to protect our employees and makes our work easier In case of noncompliance with this rule the devices are returned to the sender without att...

Страница 18: ...sswiderstand p Durchflusswiderstand p Durchflussbereich 0 bis 50 l min Flow range 0 up to 120 l min Flow range Q Flow range Q Flow range Q Flow range Q Flow range Q Flow range pressure drop p Flow ran...

Страница 19: ...sseinheit AR 2500 R Flanschanschlussma e DIN EN 1092 Gewicht 120 kg FHG1xx5 FHG1xx7 Preamplifier Preamplifier Earthing Earthing Weight 81 kg Weight 81 kg Weight 120 kg Sensor module Sensor module Conn...

Страница 20: ...ier Preamplifier Earthing Earthing on both sides on both sides M8 15 deep M8 15 deep Weight 12 kg Weight 12 7 kg Weight 24 8 kg Weight 22 kg Sensor module Test Port Test Port Sensor module Connection...

Страница 21: ...2 Evaluating unit e g display Flow sensor FHG Q 8l min V 11 53 l direction pos f 424 9 Hz control input 0 voltage blue channel 1 white channel 2 black Error Direction grey 25 Pin Assignment Channel 1...

Страница 22: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Страница 23: ...city Indicators Turbine Paddlewheel Systems Totalizers Batch Controllers pH CONDUCTIVITY pH Electrodes Testers Accessories Benchtop Laboratory Meters Controllers Calibrators Simulators Pumps Industria...

Отзывы: