Using the Cal-Lock Kit
6-1
6
The CT485B-CAL-LOCK kit that comes with the CT485B is used to preserve and
protect the settings that were entered into the CT485B during the calibration
process. The CT485B-CAL-LOCK cover shields the HIGH, SET and LOW
calibration buttons to prevent any additional button pressing. This method is
more “permanent” than using the control panel door lock. Here’s the procedure:
USE EXTREME CAUTION WHEN USING THE GLUE. READ
THE WARNING LABEL ON THE TUBE.
1.
Calibrate the CT485B.
2.
Apply a small amount of glue to the back of the cover. Refer to Figure 6-1.
Figure 6-1. Glue Application
DO NOT LET GLUE COME IN CONTACT WITH BUTTONS.
3.
Place cover over calibration buttons.
4.
Fill in the label information.
5.
Apply label to cover as shown in Figure 6-2.
Figure 6-2. Label Application
An additional label is supplied for future calibrations. A
damaged label or removal of the label implies that the
calibration of the recorder may have been altered.
NOTE
CALIBRA
TED
BY:
DATE:
CAUTION
WARNING
!