5.3 Calibrating Humidity Readings
If you have access to accurate temperature and humidity calibration equipment,
you may want to calibrate the instrument on your own as described in this
section. Figure 5-5 shows how to set up the calibration equipment. The
following flowchart shows the basic path that you should take to calibrate the
sensor.
Take 33%
Solution
Reading
(after waiting
one hour)
Take 75%
Solution
Reading
(after waiting
one hour)
Go to Temperature Calibration
(Section 5.4)
Too Low?
Too High?
G o t o
S e c t i o n
5 . 3 . 1
C a s e 2
Go to Section 5.2
to change Calibration Code
and probe label
G o t o
S e c t i o n
5 . 3 . 1
C a s e 1
Too Low?
Too High?
G o t o
S e c t i o n
5 . 3 . 2
C a s e 2
G o t o
S e c t i o n
5 . 3 . 2
C a s e 1
33% Reading
is Correct
75% Reading
is Correct
Go to Section 5.2
to change Calibration Code
and probe label
Calibrating the Recorder
5
5-4