Chilled mirror hygrometers (CMH) use a primary fundamental principal measurement which is a change in resistance. 

This principle works when a sample of process gas flows across a special polished coated mirror.  The mirror 

temperature is lowered by thermal electric coolers (located beneath the mirror block/mirror) until the water vapor of gas 

condenses out on to the mirror in the form of either dew or frost. There is an optical circuit which constantly monitors 

the mirror surface and looks for equilibrium between the optical LED detectors (a reference and the dew detector).  The 

temperature of the mirror surface is measure by a highly accurate Platinum Resistance Thermometer (PRT) which is 

embedded and sealed in the mirror block which is thermally connected to the mirror. The PRT resistance measurement is 

what makes this a fundamental dew point measurement. 

In more detail, a fundamental measurement is such that the parameter being measured, is the same as the 

Standard International definition of that parameter.  In the case of the Chilled Mirror Hygrometer, that measurement is 

the temperature of the mirror surface covered by a water layer that is in equilibrium between the liquid and gaseous 

phases of water. If that temperature measurement is accomplished by a platinum resistance thermometer, the 

measurement is equivalent to the definition of dew point. Tables constructed by NIST, notably by Wexler and 

Greenspan, establish the equivalence of dew point and the vapor pressure of water.  Using this equivalence and a second, 

fundamental, temperature measurement, realizes the definition of Relative humidity: Vapor pressure of water/saturation 

vapor pressure (temperature) of the gas being measured, X100(%) i.e. %RH =100*P

H2Ov

/P

SAT

. Addition of a 3

rd

 

measurement, the pressure of the gas being sampled, allows measurement of almost every other humidity parameter. 

Some further unit examples: absolute Humidity (g

H2Ov

/m

3

), mixing ratio (g

H2Ov

 /kg

DRY

), Parts per million (P

H2Ov

/(P

Tot

P

H20

)), enthalpy(h), to name a few. 

Chilled mirrors hygrometers are used where accuracy, typ-0.2°C or better, repeatability, stability and no 

drift are paramount. CMH are the preferred reference standard in most calibration labs. Also, CMH work well due to 

their ruggedness, their inertness, their simple maintenance and ability to accurately measure sample gasses remotely 

from process, i.e. ovens, dryers, furnaces, etc. Dew points above ambient can be measured by heating the chilled mirror 

and associated components. Chilled mirrors can be, and are, constructed as an insertion probe, provided the temperature 

of the process is < ~95 °C. 

In closing, the upside of the chilled mirror hygrometer is that it uses a fundamental principal, it can be designed to be 

inert to many corrosive gases, it is a repeatable in its measurement, there is no measurable hysteresis, it is very accurate 

and has a very long operating life. The downside is that it is expensive and is typically slower in response time when 

compared to other secondary principal humidity measurements. 

Содержание CMDP Series

Страница 1: ...1 User s Guide Shop online at omega com e mail info omega com For latest product manuals omega com en us pdf manuals Series CMDP A chilled mirror hygrometer TM...

Страница 2: ...eering Inc Headquarters 800 Connecticut Ave Suite 5N01 Norwalk CT 06854 Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Canada only Engineering Service 1 800 872 9436 USA...

Страница 3: ...3 The information contained in this document is believed to be correct but OMEGA accepts no liability for any errors it contains and reserves the right to alter specifications without notice...

Страница 4: ...le maintenance to cleaning the mirror requires using a wetted cotton swab to restore the transmitter to normal operation Because of its construction the customer can enjoy a long life of excellent per...

Страница 5: ...60C The unit also has a RS232 communication connection that can be accessed using either the Hyper terminal or TeraTerm program which is available on most computers Notes Even though the unit is scale...

Страница 6: ...ng commands Calibrating Current Outputs There are 2 methods for calibrating the 4 20 mA outputs Method 1 uses a digital ammeter capable of 1 uA Resolution The second method uses high precision 005 Res...

Страница 7: ...one of these three states a The unit was just powered on and is going through its ARC cycle for 3 5minutes b The mirror is dirty has contaminants on it and it needs to be cleaned Refer to the clean m...

Страница 8: ...be removed Once removed twist the outer shell to the correct hole for airflow and then re install the set screw to hold the outer shell in place For example if the larger airflow hole has been select...

Страница 9: ...Female NPT Class 125 The sensor head black flow rate cap should be installed at least 1 into the process gas If the reading is oscillating you might need to adjust the black flow rate cap slowing the...

Страница 10: ...INTERPRET GEOMETRIC TOLERANCING PER DRAWN CHECKED ENG APPR MFG APPR Q A COMMENTS NAME DATE TITLE A CM 10 D THE INFORMATION CONTAINED IN THIS DRAWING IS THE SOLE PROPERTY OF INSERT COMPANY NAME HERE A...

Страница 11: ...ler and Greenspan establish the equivalence of dew point and the vapor pressure of water Using this equivalence and a second fundamental temperature measurement realizes the definition of Relative hum...

Страница 12: ...9...

Страница 13: ...therwise shall not exceed the purchase price of the component upon which liability is based In no event shall OMEGA be liable for consequential incidental or special damages CONDITIONS Equipment sold...

Страница 14: ...Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U Industrial pH...

Отзывы: