Section 1 - Introduction

Your BB700 Series Blackbody Calibration Source has been designed for ease of 

use and reliability whenever you have the need to test or calibrate non-contact 

infrared temperature instruments. It is important that you read this manual 

completely and follow all safety precautions before operating this instrument.

1.1 Precautions

• Follow all safety precautions and operating instructions outlined in this 

manual.

• Never leave your calibrator unattended when in use.
• Keep out of reach of children.
• Never touch the aperture plate or header cavity when hot.
• Never place any flammable or explosive material near the cavity opening.
• Do not operate in flammable or explosive environments.
• Never operate with a power cord other than the one provided with your unit.
• Remove or disconnect main power cord before attempting any maintenance or 

fuse replacement.

• Do not connect this unit to a non-grounded, non-polarized outlet or power 

source.

• This unit is intended for indoor use only. Avoid exposure to any moisture or 

rain.

• Do not connect the serial ports to equipment with exposed, hazardous, live 

voltages.

• The operator of this instrument is advised that if the equipment is used in 

a manner not specified in this manual, the protection provided by the 

equipment may be impaired.

There are no user serviceable parts inside your unit. Attempting to repair or service your unit may 

void your warranty.

1.2 

Safety Warnings and IEC Symbols

This device is marked with international safety and hazard symbols in 

accordance with IEC 1010. It is important to read and follow all precautions and 

instructions in this manual before operating or commissioning this device as it 

contains important information relating to safety and EMC. Failure to follow all 

safety precautions may result in injury and or damage to your calibrator. Use of 

this device in a manner not specified by the manufacturer may impair protection 

provided within the unit.

Introduction

1

1-1

NOTE:

Содержание BB700 Series

Страница 1: ...latest product manuals www omegamanual info BB705 Very High Temperature Laboratory Blackbody Calibration Source Shop online at omega com User sGuide TM This datasheet has been downloaded from http www...

Страница 2: ...pecifications without notice omega com info omega com Servicing North America U S A Omega Engineering Inc Headquarters Toll Free 1 800 826 6342 USA Canada only Customer Service 1 800 622 2378 USA Cana...

Страница 3: ...4 Controller Settings 3 5 3 4 1 Changing the Temperature Setpoint 3 5 3 4 2 Changing the Controller Parameter Settings 3 6 3 5 Temperature Reference Meter Settings 3 7 3 6 Target Cavity Temperature T...

Страница 4: ...e 3 3 5 Rear Panel 3 4 6 Controller Default Parameters 3 6 7 Controller Programming Procedures 3 7 8 Approximate Target Plate Temperature Transition Times 3 8 9 Connecting a PC to the BB705 s Controll...

Страница 5: ...t this unit to a non grounded non polarized outlet or power source This unit is intended for indoor use only Avoid exposure to any moisture or rain Do not connect the serial ports to equipment with ex...

Страница 6: ...The calibration source with built in PID digital controller maintains an accuracy of 0 25 of reading 1 C The cavity has a 44 5 mm 1 75 diameter aperture with an emissivity of 0 99 and can be set to a...

Страница 7: ...reshipment is necessary The following items are supplied in the box Item OMEGA Part Number BB705 Blackbody Calibration Source BB705 or BB705 230 User Manual M 3638 Calibration Certificate CAL 3 3 DATA...

Страница 8: ...0 VAC 50 60 Hz Models On 230 VAC 50 60Hz models an International style power cord with the proper color code and approvals is provided with stripped wire ends for connection to the proper connector us...

Страница 9: ...3 3 1 C A U T I O N CAVITY MAY B E H O T 1000C 1000 00C 1000C SETPTS MAX DEV MENU TEMPERATURE REFERENCE METER TEMPERATURE CONTROLLER SET VARIABLE PROCESS VARIABLE CAVITY APERTURE APERTURE PLATE POWER...

Страница 10: ...is used to set the setpoint temperature and change engineering units Temperature Reference Meter The temperature reference meter displays the cavity temperature with high precision See Section 3 5 for...

Страница 11: ...ronment only Never operate the unit outside or around children Nothing should come in contact with the heater cavity aperture or aperture plate Even when the unit is off Never unplug the unit while it...

Страница 12: ...E METER CAUTION FOR CONTINUED PROTECTION AGAINST THE RISK OF FIRE REPLACE WITH ONLY THE SAME TYPE RATING AND SIZE OF FUSE DISCONNECT POWER FROM SOURCE BEFORE CHANGING FUSE INSTALLATION CATEGORY II 208...

Страница 13: ...t temperature of the target plate Setpoint Temperature This field displays the desired target plate temperature Once the target plate reaches this desired temperature both displays will read the same...

Страница 14: ...hrough LED indicators An indicator may be lit continuously blink or shut off entirely to indicate that the process is on being pulsed or off respectively 3 4 2 Changing the Controller Parameter Settin...

Страница 15: ...ress the MENU key until the display reads RDG CNF 2 Press the MIN key until the display reads RDG 5 3 Set this parameter as desired using the MAX key 0 C 1 F 2 K 4 Press MENU to store the settings 5 P...

Страница 16: ...to another First determine the current heater cavity temperature and go to the corresponding row Now decide the new heater cavity temperature and go to the appropriate column Note that there are X s a...

Страница 17: ...ers should be viewed or queried from the controller directly It is highly recommended that baud rate for the controller be modified on the controller directly Note that both the BB705 and the computer...

Страница 18: ...case R To read a parameter from the controller M To temporarily modify a controller parameter lost upon shutdown E To modify a controller param in non volatile mem saved even after shutdown parameter...

Страница 19: ...ing soot from heater cavity be sure to wear protective goggles NEVER REMOVE SOOT FROM A HOT OR WARM CAVITY Be sure unit has cooled to room temperature and is disconnected from power source before vacu...

Страница 20: ...ontrol Sensor Platinum Thermocouples Accuracy 1 C 0 25 rdg 1 8 F 0 25 rdg Ref Meter Display Accuracy 0 06 C over entire range Resolution 0 01 on temperature reference meter Overall System Stability 0...

Страница 21: ...ler display shows Error a Unit requires service contact our and the target plate will not get customer service department hot or cold 4 Target plate temperature will a Verify that the controller is se...

Страница 22: ...e ratio of energy emitted by a surface to the energy emitted by a blackbody at the same temperature IEC International Electrotechnical Commission Infrared IR A range of the electromagnetic spectrum ex...

Страница 23: ...ource Calibration Range 32 to 400 C 90 to 752 F Accuracy 1 4 C 2 5 F Emissivity 0 95 Ambient Temp 0 to 40 C 32 to 104 F Cavity Size 28 6 mm 1 125 in Power 115 230V 50 60 Hz 175W BB704 4 Target Plate B...

Страница 24: ...NOTES BB705 Very High Temperature Laboratory Blackbody Calibration Source 9 2...

Страница 25: ...A if the unit is found to be defective it will be repaired or replaced at no charge OMEGA s WARRANTY does not apply to defects resulting from any action of the purchaser including but not limited to m...

Страница 26: ...s M U Turbine Paddlewheel Systems M U Totalizers Batch Controllers pH CONDUCTIVITY M U pH Electrodes Testers Accessories M U Benchtop Laboratory Meters M U Controllers Calibrators Simulators Pumps M U...

Отзывы: