Section 6 - Specifications
Heater Cavity Temperature Range: 100°C to 1046°C
(212°F to 1915°F)
Ambient Environmental Conditions:
Temperature:
0°C to 35°C (32°F to 95°F)
Humidity:
0 to 90% RH, non-condensing
Power:
BB705
115 VAC~, 50/60 Hz, 1100W
BB705-230VAC
230 VAC~, 50/60 Hz, 1100W
Internal Control Sensor:
Platinum Thermocouples
Accuracy:
±1°C, ±0.25% rdg
(±1.8°F, ±0.25% rdg)
Ref. Meter Display Accuracy:
±0.06°C (over entire range)
Resolution:
0.01° (on temperature reference meter)
Overall System Stability:
±0.4°C per 8-hr. period
Target Plate Emissivity: 0.99
Cavity Aperture Diameter:
44.4 mm (1.75")
Heat-Up Time:
80 minutes from 100°C to 1000°C
(from 212°F to 1832°F)
Dimensions:
566 W x 520 H x 600 D
(22.3 W x 20.5 H x 23.6" D)
Weight:
31 kg. (68 lb.)
Installation Category II
Specifications
6
6-1