Specifications
8
8-4
LASER SIGHTING
Wavelength (Color):
630-670 nanometers (red)
Operating Distance:
Laser Dot
2 to 40 ft.
Laser Circle
2 to 15 ft.
Max. Output Optical Power: <1mW at 75°F ambient temperature,
Class II Laser Product
European Classification:
Class 2, EN60825-1
Maximum Operating Current: 25mA at 5.5 V
FDA Classification:
Complies with 21 CFR Chapter 1, Subchapter J
Beam Diameter:
5 mm
Beam Divergence:
<2mrad
Laser Configuration:
Dot/Circle switchable except for OS53x-CF
and OS523E-3 models
Laser Status:
Flashing or continuous, set via keypad
Operating Temperature:
0°C to 50°C (32°F to 122°F)
Operating Relative Humidity: 95% or less without condensation
Power Switch:
Slide switch, ON - OFF
Power Indicator:
Red LED
Power:
Supplied by the thermometer
Identification Label:
Located on the bottom of the thermometer
Warning & Certification Label: Located on the left side of the thermometer
(for the label layout, refer to the inside back cover)
DISTANCE MEASURING (Model HH-DM)
Range:
0.9 to 9 m (3' to 30')
Accuracy:
1% of Rdg or 3 cm (0.1') whichever is greater
Units of Measure:
Meter or Feet – switchable via C button
Sensor:
Ultrasonic transducer
Power:
9 Volts Battery
Battery Life:
50 hours, continuous
Operating ambient
0 to 40°C
temperature:
Operating relative humidity: 30 to 70% RH
Auto power shut off:
Approx. 7 minutes after the last key press
Содержание OMEGASCOPE OS523E
Страница 3: ...Unpacking Instructions n4 Notes...
Страница 5: ...ii...
Страница 15: ...General Description 1 1 8 Notes...
Страница 61: ...7 4 Troubleshooting Guide 7 Notes...
Страница 67: ...Specifications 8 8 6 Notes...
Страница 69: ...Glossary of Key Strokes 9 9 2 Notes...
Страница 79: ...Appendix Determining an Unknown Emissivity C C 4 Notes...