Specifications
8
(Specifications are for all models except where noted)
THERMOMETER
Measuring
OS530HRE,
-30°C to 121°C (-22°F to 250°F)
Temperature
OS530LE,
Range:
OS533E,OS532E:
-23°C to 538°C (-10°F to 1000°F)
OS534E
-23°C to 871°C (-10°F to 1600°F)
OS523E
-18°C to 1371°C (0°F to 2500°F)
OS524E
538°C to 2482°C (1000°F to 4500°F)
±1% of reading or
3°F whichever is greater
(2% Rdg for temp > 2000°F for OS524E)
Field of Vision:
OS532E, OS530LE 10:1
OS533E, OS530HRE 20:1
OS534E
30:1
OS53x-CF
0.15"@6"
OS523E-1
30:1
OS523E-2
60:1
OS523E-3
68:1
OS524E
110:1
Repeatability:
± (1% rdg + 1 digit)
Resolution:
1°C or 1°F (0.1°C or °F for OS530HRE)
Response Time:
100 msec
Spectral Response:
8 to 14 microns (2 to 2.5 microns, OS524)
Thermocouple Input:
Type K, -18 to 871°C (0 to 1600°F)
(OS532E, OS533E, OS534E only)
Input Connection:
SMP Connector
Thermocouple Display
Accuracy @ 24°C (75°F)
Ambient Temperature:
±3°C (±5°F)
Thermocouple Display
Response Time:
2 seconds
Operating Ambient
0°C to 50°C
Temperature:
(32°F to 122°F)
Operating Relative
Humidity:
95% or less without condensation
Display:
Backlit LCD dual display
Keypad:
4 position, tactile feed-back membrane switch
8-1
Accuracy (24°C or 75°F
Ambient Temperature and
at emissivity of 0.95 or
greater):
Содержание OMEGASCOPE OS523E
Страница 3: ...Unpacking Instructions n4 Notes...
Страница 5: ...ii...
Страница 15: ...General Description 1 1 8 Notes...
Страница 61: ...7 4 Troubleshooting Guide 7 Notes...
Страница 67: ...Specifications 8 8 6 Notes...
Страница 69: ...Glossary of Key Strokes 9 9 2 Notes...
Страница 79: ...Appendix Determining an Unknown Emissivity C C 4 Notes...