VA Series Vapor Actuated Dial Thermometer

Vapor actuated thermometers give excellent readings when the bulb
temperature is either above or below the temperature of the indicating
head. Avoid applications in which the bulb and head are the same
temperature (ambient applications). If the bulb “crosses ambient” (i.e.,
goes from hotter than the indicating head to cooler or vice versa), some
pointer fluctuations will occur for up to two minutes during this
transition.

Recalibration

The VA Series vapor actuated thermometers are manufactured with
threaded, press fit and snap-ring lenses which can be easily removed
for field calibration.
To recalibrate your vapor actuated thermometer:
1. Remove the thermometer from its fitting.
2. Remove the lens.
4. Immerse the bulb totally in an agitated liquid for at least 3 minutes

with a master reference thermometer, until they reach equilibrium.

5. Use a wrench, coin or screwdriver to turn the slotted hex nut on the

back of the thermometer case until the pointer is at the proper
reading on the scale.

6. Replace the lens and reinstall the thermometer in  its fitting.

Care of the Vapor Actuated Thermometer

• Avoid bending the stem, as this will result in frictional errors in

indication.

• Do not kink or flatten the capillary tube.
• Dampen any capillary vibration, as this may cause capillary breakage

over time.

6

Содержание M1982/0902

Страница 1: ...capillary tube length Available Models Front Flange U Clamp Panel Mount Rear Flange Panel Mount Surface Mounted Figure 1 Case Style Installing the Vapor Actuated Thermometer Unpacking Instructions Rem...

Страница 2: ...ssary Case and Mounting Dimensions Front Flange U Clamp Rear Flange Panel Mount Panel Mount Surface Mount Figure 2 Case and Mounting Dimensions Installing the Indicating Head Surface Mount Rear Flange...

Страница 3: ...he front of the panel pass the bulb and capillary through the cutout 4 Affix the head to the panel 4a U clamp Model Tighten U clamp s against the back of your panel 4b Front Flange Model Screw the fla...

Страница 4: ...nections Tighten the thermometer to its fitting using a wrench applied to the hex or flats above the threads Never tighten by hand 1 Unscrew and remove the 1 2 NPT fitting using a 7 8 wrench 2 Screw t...

Страница 5: ...fitting with an 11 16 wrench Bendable Extensions I Place the bulb in the thermowell or directly in the process Use a thermowell whenever the bulb would be exposed to pressure fluid velocity or corros...

Страница 6: ...ing lenses which can be easily removed for field calibration To recalibrate your vapor actuated thermometer 1 Remove the thermometer from its fitting 2 Remove the lens 4 Immerse the bulb totally in an...

Страница 7: ...teel or nickel plated copper Bulb O D 7 16 optional 1 4 O D x 2 length on 40 to 65 F range only Bulb lengths vary from 2 to 6 depending on capillary length Capillary Nickel plated copper copper with b...

Страница 8: ...medical application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hol...

Отзывы: