VA Series Vapor Actuated Dial Thermometer

Use a thermowell in all pressurized applications. Be sure
case temperature does not exceed 200°F (93°C).

Thermowells

1. Remove the thermowell and screw it into the process so that no

leaking occurs. Use thread sealant as required.

2. Place the bulb in the thermowell. Tighten the bulb in place by

screwing the jam nut into the top of the 

1

2

" fitting with an 

11

16

" wrench.

Bendable Extensions

I . Place the bulb in the thermowell, or directly in the process.

Use a thermowell whenever the bulb would be exposed to
pressure, fluid velocity, or corrosive media.

2. Slide the compression fitting down and screw the 

1

2

" NPT fitting into

the process (or thermowell) threads.

3. Tighten the top hex of the sliding union until the bendable extension

does not slide in or out.

4. Gently bend the remaining portion of the extension out of the way.

Operation

Recommended operating temperature is in the upper 

2

3

of the dial scale.

N O T E

N O T E

5

Содержание M1982/0902

Страница 1: ...capillary tube length Available Models Front Flange U Clamp Panel Mount Rear Flange Panel Mount Surface Mounted Figure 1 Case Style Installing the Vapor Actuated Thermometer Unpacking Instructions Rem...

Страница 2: ...ssary Case and Mounting Dimensions Front Flange U Clamp Rear Flange Panel Mount Panel Mount Surface Mount Figure 2 Case and Mounting Dimensions Installing the Indicating Head Surface Mount Rear Flange...

Страница 3: ...he front of the panel pass the bulb and capillary through the cutout 4 Affix the head to the panel 4a U clamp Model Tighten U clamp s against the back of your panel 4b Front Flange Model Screw the fla...

Страница 4: ...nections Tighten the thermometer to its fitting using a wrench applied to the hex or flats above the threads Never tighten by hand 1 Unscrew and remove the 1 2 NPT fitting using a 7 8 wrench 2 Screw t...

Страница 5: ...fitting with an 11 16 wrench Bendable Extensions I Place the bulb in the thermowell or directly in the process Use a thermowell whenever the bulb would be exposed to pressure fluid velocity or corros...

Страница 6: ...ing lenses which can be easily removed for field calibration To recalibrate your vapor actuated thermometer 1 Remove the thermometer from its fitting 2 Remove the lens 4 Immerse the bulb totally in an...

Страница 7: ...teel or nickel plated copper Bulb O D 7 16 optional 1 4 O D x 2 length on 40 to 65 F range only Bulb lengths vary from 2 to 6 depending on capillary length Capillary Nickel plated copper copper with b...

Страница 8: ...medical application used on humans or misused in any way OMEGA assumes no responsibility as set forth in our basic WARRANTY DISCLAIMER language and additionally purchaser will indemnify OMEGA and hol...

Отзывы: