39
Variables Table
Category
Variable Name
Description
Get
Set
Calibration_Name
Name of calibration
V
Calibration_Date
Calibration date (MMDDYY).
V
Calibration_Time
Calibration time
(HHMM, e.g., 1545).
V
Unit
Weighing unit:
ton, kg, g, lb or oz.
V
Resolution
Weighing resolution:
0.0001, 0.0002, 0.0005, 0.001,
0.002, 0.005,
0.01, 0.02, 0.05,
0.1, 0.2, 0.5,
1, 2, 5,
10, 20 or 50.
V
Full_Capacity
Full capacity of the load cell.
V
Maximum_Load
Maximum applied load on the
load cell.
V
Calibration
Info
Load_Cell_Output
Output of the load cell:
0 = Unknown
1, 2 or 3 = 1, 2 or 3 mV/V.
V
Filter1
V
*
Filter2
V
*
Filtering
Decimator
There are
two
filters. The
first filter
is
basically a 1st level moving average
filter of size
Filter1
(2-256). Then
depending on
Decimator
(1-1000),
every Decimator-th result from the 1st
level filter is put through another
moving average filter of size
Filter2
(2-256), which is the
second filter
.
*
S
etting
these three variables causes
a
board reset
, so it is a
slow
function.
Therefore, in order to set
all of them
there is one more option: the method
Set_Filtering , which works faster
than three individual activations of
the Set_Variable method.
V
*
Содержание High Speed Load Cell Interface Card...
Страница 51: ...50 A 2 Connecting External Devices to the Inputs...
Страница 52: ...51 A 3 LCIC WIM I O Analog Out...
Страница 55: ...54 Appendix C Load Cell Connections...