www.microscopenet.com

 

 

- 7 -

3.

 

Specifications 

 

General 

Objective Tube Length 

160mm 

Viewing Head 

Binocular, inclined 30º, swiveling 360º, interpupillary distance 
55-75mm, adjustable diopter   

Eyepieces WF10X, 

P16X 

Nosepiece Quadruple 

Objectives 

Achromatic 4X, 10X, 40X(spring), 100X(spring, oil) 

Focus 

system 

Coaxial coarse and fine focusing, minimum fine focusing 
adjustment at 0.002mm, range 28mm 

Condenser Abbe, 

NA=1.25 

Stage 

Double layer mechanical stage, area 140mmx140mm, 
movement range 75mmx50mm 

Illumination 

Halogen lamp 6V/20W 

 

Eyepieces 

Designation 

Magnification 

Field of View 

Focal Length 

Wide Field 

10X 

18mm 

24.94mm 

Plan Field 

16X 

11mm 

15.58mm 

 

Objectives

 

   

Magnification 

Numerical 

Aperture 

Thickness of 

slide cover 

Focus 

Working 

Distance 

Working 

Mode 

4X 

0.10 

0.17mm 31.04mm 37.5mm  Dry 

10X 0.25 

0.17mm 

17.13mm 

7.316mm 

Dry 

40X(S) 0.65 

0.17mm 

4.65mm 

0.632mm 

Dry 

100X(S, Oil) 

1.25 

0.17mm 

2.906mm 

0.198mm 

Oil 

 

Magnifications

 

   

 

Eyepiece  10X 16X 10X 16X 10X 16X 10X 16X 

Objective 4X 

10X 

40X  100X 

Magnification  40X  64X  100X 160X 400X 640X 1000X 

1600X 

 

Содержание M829

Страница 1: ...User Manual Binocular Biological Microscope Model M829 MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Installation 3 2 Operation 5 3 Specifications 7 4 Troubleshooting Guide 8...

Страница 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...ently for dirt or oil clean if necessary Use the 10x objective to initially focus the image of your specimen When changing the objective magnification rotate the objective nosepiece until you hear a c...

Страница 6: ...the thumb screw 1 7 Installing the mirror optional your model may not have one 1 Turn the power off and unplug the power cord 2 Screw off the light cover on the base 3 Thread the black disc onto the...

Страница 7: ...condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives Note The centering of the c...

Страница 8: ...sharp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 2 8 Adjusting tension knob The tightness of t...

Страница 9: ...e layer mechanical stage area 140mmx140mm movement range 75mmx50mm Illumination Halogen lamp 6V 20W Eyepieces Designation Magnification Field of View Focal Length Wide Field 10X 18mm 24 94mm Plan Fiel...

Страница 10: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Страница 11: ...sity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...