2.

 

Operation 

2.1 Adjusting illumination 

1)  Connect the power cord and turn on the main switch.   
2)  Rotate the variable intensity dial to increase or decrease the brightness. 

2.2 Placing specimen 

1)  Place the slide on the mechanical stage.   
2)  Use the slide holder to gently secure the slide. 
3)  Turn the X and Y translational adjustment knobs to position the specimen for viewing. 

Caution:

 Be sure not to allow an objective to touch a specimen slide when changing 

objectives. 

2.3 Focusing 

1)  With the 10x objective in position, raise the mechanical stage using the coarse focus 

knob until the specimen is close to the objective. 

2)  Turn the coarse focus knob until the specimen is in focus.   
3)  Then use the fine focus knob to obtain a sharp image.   
4)  You may now switch to another magnification objective. 

Tips:

 

To prevent your specimen slide from making contact with an objective, raise the 

stage to its highest position without contacting the 100x objective, then tighten the upper 
limit mechanical stage lever.

 

2.4 Adjusting condenser 

1)  Turn the condenser focus knob to raise or lower the condenser.   
2)  The condenser is raised when using high magnification objectives and lowered when 

using low magnification objectives. 

Note:  

  The centering of the condenser and the light axis of the objective are factory 

adjusted. Do not attempt re-adjustment. 

  The highest position of the condenser has been factory adjusted. Do not attempt 

re-adjustment.

 

2.5  Adjusting iris diaphragm 

Move the Iris Diaphragm Lever left or right to adjust the aperture size.   

Note: 

The iris diaphragm is designed to adjust the aperture size, not to adjust 

brightness. Generally, opening the diaphragm to 70-80% of the N.A. value of the 
respective objective will provide an image of acceptable quality. If you want to observe 
the image of the iris diaphragm, remove one eyepiece and look through the tube. You 
will see a dark circle encroaching on the bottom of the tube. 

 

2.6  Adjusting interpupillary distance 

While observing with both eyes, hold the left and right eye tubes then slowly slide the 
tubes in and out. The interpupillary distance is correct when the left and right fields of 
view converge completely into one image. 

2.7  Adjusting eyepiece diopter 

1)  Using the 10x objective and your right eye only, observe your specimen through the 

Содержание M829

Страница 1: ...User Manual Binocular Biological Microscope Model M829 MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Installation 3 2 Operation 5 3 Specifications 7 4 Troubleshooting Guide 8...

Страница 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...ently for dirt or oil clean if necessary Use the 10x objective to initially focus the image of your specimen When changing the objective magnification rotate the objective nosepiece until you hear a c...

Страница 6: ...the thumb screw 1 7 Installing the mirror optional your model may not have one 1 Turn the power off and unplug the power cord 2 Screw off the light cover on the base 3 Thread the black disc onto the...

Страница 7: ...condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives Note The centering of the c...

Страница 8: ...sharp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 2 8 Adjusting tension knob The tightness of t...

Страница 9: ...e layer mechanical stage area 140mmx140mm movement range 75mmx50mm Illumination Halogen lamp 6V 20W Eyepieces Designation Magnification Field of View Focal Length Wide Field 10X 18mm 24 94mm Plan Fiel...

Страница 10: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Страница 11: ...sity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...