4.

 

Troubleshooting Guide 

 

OPTICAL PROBLEMS

 

Problem

 

Cause

 

Solution

 

Darkness at the 
periphery or uneven 
brightness in the 
field of view 

Revolving nosepiece not in click 
stop position 

Revolve the nosepiece to click-stop 
position by swinging the objective 
correctly into the optical path 

The light source of the bulb is not 
at the center 

Adjust the position of the bulb 

Dirt or dust on the 
view 

Dirt or dust on the lens eyepiece, 
condenser, objective, collector 
lens or specimen 

Clean the lens with a camera 
cleaning kit 

Poor image quality 

No slide cover attached to the 
slide 

Attach a 0.17mm slide cover 

Slide cover is too thick or thin 

Use a slide cover of the appropriate 
thickness (0.17mm) 

Slide may be upside down 

Turn slide over so the cover-glass 
faces up 

Immersion oil is on a dry objective 
(especially the 40x) 

Check the objectives, clean if 
necessary 

No immersion oil used with 100x 
objective 

Use immersion oil 

Air bubbles in immersion oil 

Remove bubbles 

Condenser aperture is closed or 
open too much 

Open or close properly 

Condenser is positioned too low 

Position the condenser upward 

 

 

ELECTRICAL PROBLEMS 

Problem

 

Cause

 

Solution

 

Lamp does not light 
when switched on 

No electrical power   

Check power cord connection   

Lamp bulb burnt out   

Replace bulb   

Fuse blown out 

Replace fuse 

 

 

Содержание M829

Страница 1: ...User Manual Binocular Biological Microscope Model M829 MicroscopeNet com...

Страница 2: ...Table of Contents i Caution 1 ii Care and Maintenance 2 1 Installation 3 2 Operation 5 3 Specifications 7 4 Troubleshooting Guide 8...

Страница 3: ...ve or any other component disconnect the power cord immediately and wipe up the spillage Otherwise the instrument may be damaged 5 Important the lamp lamp housing and adjacent parts will become very h...

Страница 4: ...he optical surface use a soft cloth or cotton swab dampened with a lens cleaning solution available at camera stores All optical lenses should be swabbed using a circular motion A small amount of abso...

Страница 5: ...ently for dirt or oil clean if necessary Use the 10x objective to initially focus the image of your specimen When changing the objective magnification rotate the objective nosepiece until you hear a c...

Страница 6: ...the thumb screw 1 7 Installing the mirror optional your model may not have one 1 Turn the power off and unplug the power cord 2 Screw off the light cover on the base 3 Thread the black disc onto the...

Страница 7: ...condenser focus knob to raise or lower the condenser 2 The condenser is raised when using high magnification objectives and lowered when using low magnification objectives Note The centering of the c...

Страница 8: ...sharp image is obtained 3 Since both sides are adjustable you may also do the above in the opposite way in other words left eye first and right eye second 2 8 Adjusting tension knob The tightness of t...

Страница 9: ...e layer mechanical stage area 140mmx140mm movement range 75mmx50mm Illumination Halogen lamp 6V 20W Eyepieces Designation Magnification Field of View Focal Length Wide Field 10X 18mm 24 94mm Plan Fiel...

Страница 10: ...quality No slide cover attached to the slide Attach a 0 17mm slide cover Slide cover is too thick or thin Use a slide cover of the appropriate thickness 0 17mm Slide may be upside down Turn slide over...

Страница 11: ...sity control dial Aperture diaphragm closed too far Open to the proper setting Condenser position too low Position the condenser upward MECHANICAL PROBLEMS Problem Cause Solution Image will not focus...

Отзывы: