Acquiring and saving images
21
5)
Click the
Unit...
button.
"
Select
m
(for meters) in the
Basic unit
list in the
Set Unit
dialog box.
"
Select, e.g.,
µ
in the
Scale
list if you wish the calibration length to be
shown in µm. Confirm by clicking
OK
.
6)
Enter this magnification level into the
Magnification
field.
7)
Enter the length you are using to calibrate with into the
Calibration length
field.
"
The distance between the scale marks 30 and 80 in the example shown
is 500 µm.
8)
Select the
Arbitrary
option in the
Calibration
group.
9)
Now click the
Calibrate
button.
"
The mouse pointer will appear within the active image.
10)
Position the mouse pointer at the starting point of the calibration length and
leftclick.
11)
Then position the mouse pointer at the final point of the calibration length
and confirm by leftclicking. The blue line must be the exact same length as
the calibration length you entered in the
Calibrate Image
dialog box.
"
The calibration data for the selected magnification are now shown in the
X calibration
and
Y calibration
fields.