MV-1/MV-2 Endoscope Disinfectors
Frequently
Asked
Questions
09FQ0015 5/18
Two Corporate Center Drive, Melville, NY 11747
Copyright 2001, Olympus America Inc.
Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part
thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do
Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.
Cabling and Connectivity
Connection of the MV-1/MV-2 to ancillary equipment.
See MV-1/MV-2 Endoscope Disinfectors User/Service Manual
See also document #09-WI-0239, MV-1/MV-2 Endoscope Disinfectors Installation Procedure
Note:
The pre-installation instructions in paragraphs 8.0 – 8.5 of document
MUST
be completed
before performing the following connections. Also refer to
Disinfectors User/Service Manual.
1.
Connection of the Drain to the MV-1/MV-2
#
Flexible drain hose
➨
Connect the flexible drain hose to the white drain fitting on the back of the MV-1/MV-2, and
attach the screw clamp.
➨
Caution:
The drain must be 12 in. below
the drain fitting on the disinfector, and loops in the drain
hose must be eliminated to ensure that fluids drain thoroughly. Incomplete draining will result in
dilution of the LCG and/or inadequate rinsing of the endoscope.
2.
Connection of the LCG Reservoir to the MV-1/MV-2
#
LCG Reservoir
➨
The LCG reservoir can be installed either under the counter top, or on the bottom of the mobile
cart.
➨
Caution:
The LCG reservoir must be placed below the MV-1/MV-2 basin. Incorrect placement of
the reservoir may cause the LCG to seep back into the MV-1/MV-2 basin.
➨
Connect the cut ends (without fittings) of the LCG reservoir tubes (Item CLM-1000) to the barbed
fittings on the LCG reservoir. Each tube can be connected to either fitting.
➨
Attach the screw clamps provided.
➨
Connect the LCG reservoir tubing ends with right angle male quick-release connectors to the
quick-release connectors labeled
Solution/In
and
Solution/Out
on the back of the MV-1/MV-2.
Each quick release connector on the tube end can be connected to either quick-release connector
on the MV-1/MV-2.
3. Connection of the Power Cord to the MV-1/MV-2
#
Power Cord
➨