MV-1/MV-2 Endoscope Disinfectors
Frequently
Asked
Questions
09FQ0015 15/18
Two Corporate Center Drive, Melville, NY 11747
Copyright 2001, Olympus America Inc.
Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part
thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do
Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.
Troubleshooting: MV-1/MV-2 Stopped in Mid-Cycle
1. Has the MV-1/MV-2 stopped in mid-cycle?
#
There could be a fluctuation in the power source voltage. If this is suspected, contact the facility’s
maintenance department.
#
Check the power cord to be sure it is securely connected.
#
The MV-1/MV-2 rinse alarm could be activated. Refer to question #1.
Troubleshooting: LCG
1. Was the LCG lost through the overflow drain?
#
Check to see if the MV-1/MV-2 is level. If necessary, adjust the legs of the MV-1/MV-2.
#
Make sure any material or debris is not covering the float tower.
#
Make sure that the float sensor is not defective. Replace the float sensor if necessary.
2. Is the LCG not completely filling the basin?
#
Make sure the LCG tubing does not contain any kinks or bends.
#
Ensure the LCG reservoir is filled to at least the minimum fill level.
#
Check the LCG-In quick connectors to be sure that they are not worn or not fully inserted. If necessary,
replace or reconnect the connectors.
#
Check to see if the LCG is leaking through the drain. If so, contact Olympus.
#
Check the operation of the OSC pumps. Replace one or more of the pumps if necessary.
#
Make sure the float sensor is not defective. If necessary, replace the float sensor.
3. Is the LCG alarm activated?
#
Refer to the possible solutions in question #2.
4. Is the LCG reservoir overfilling or is water left in the basin after the rinse cycle?
#
Check the drain hose to make sure it is below the level of the drain connection on the back of the MV-
1/MV-2.
#
Make sure the drain hose is not kinked
.
#
Ensure that the drain screen is clear.