MV-1/MV-2 Endoscope Disinfectors
Frequently
Asked
Questions
09FQ0015 11/18
Two Corporate Center Drive, Melville, NY 11747
Copyright 2001, Olympus America Inc.
Copyright 2001 Olympus America Inc. This document and the contents contained herewith remain the property of Olympus America Inc. Reproduction of this document or any part
thereof by any means (electronic or otherwise) without the expressed written consent of Olympus America Inc. is strictly prohibited. Only printed copies with "Controlled Document -- Do
Not Copy" stamped in red are considered controlled copies. All other printed copies are considered Uncontrolled and may not contain the latest document revisions.
Preventive Maintenance (Cont’d)
Refer to
Chapter 5 in the MV-1/MV-2 Endoscope Disinfectors User/Service Manual.
Also, refer to
document #
09-WI-0240, MV-1/MV-2 Endoscope Disinfectors Confidence Test.
In order to ensure correct operation of the MV-1/MV-2, the following quality assurance procedures must be
performed on- the MV-1/MV-2 EACH day prior to the first cycle.
Contact Olympus if these procedures fail to yield the expected results.
Operational Phase
Area Of Concern
Recommendation
Before Operating
MV-1/MV-2 Water
Flow
A channel connector should be connected to each port in the
MV-1/MV-2 basin.
•
Turn the MV-1/MV-2 power switch
ON.
The power switch
is located on the right side of the MV-1/MV-2.
•
Press
START
on the MV-1/MV-2 control panel.
•
Wait for the pump to feed water through the channel
connectors.
•
Confirm that water is flowing properly through each
connector.
•
Note:
Flow volume will vary depending upon the type of
channel connector used. The water flow through the male
luer lock connector should be approximately 100 cc (4
oz.) in less than 15 seconds.
•
If the water flow through any connector appears to be
inadequate, check it again using a male luer lock
connector. This will determine if the inadequate water
flow is caused by the connector or the MV-1/MV-2.
•
Note:
Inadequate water flow may be caused by a
clogged valve or a malfunction of the main pump.
•
Press
STOP
on the MV-1/MV-2 control panel.
•
Note:
Contact Olympus if water does not flow properly
through all of the channel connectors.
Before Operating
MV-1/MV-2 Air
Flow
•
Press
MANUAL, DRAIN,
and then
START
on the MV-
1/MV-2 control panel.
•
While holding all tubing in the MV-1/MV-2 basin under
water, confirm that there is vigorous bubbling in all tubing.
•
Completely drain the water from the MV-1/MV-2 basin.
•
Press
STOP
on the MV-1/MV-2 control panel.
•
Note:
Contact Olympus if air does not flow properly
through ALL tubing.