DMTA017-01EN, Rev. F, August 2016
Chapter 4
30
Figure 4
‑
1 Removing the wheel screw
2.
Lift the scanner and pull the wheel down and outward.
3.
Insert the new wheel in the frame and secure it with the screw.
Make sure the bearings are placed in the correct position.
4.
Repeat the procedure for the other wheels.
To avoid the risk of injury from magnetic attraction forces between wheels, only
remove and replace one wheel at a time (always reinstall the magnetic wheel on the
scanner before removing the next one).
4.2
Replacing a Conventional Ultrasonic Probe
This section describes the procedure for removing a conventional ultrasonic probe
and wedge from the holder on the scanner.