Chapter 5 Troubleshooting
73
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Flexibility adjustment (for CF-Q160AL/I, PCF-160AL/I only)
Auxiliary water feeding (for endoscopes with auxiliary
water feeding only)
Endo-therapy accessories
Others
Irregularity
description
Possible cause
Solution
Too difficult to turn the
flexibility adjustment
ring.
The insertion tube is
looped.
Straighten the insertion tube.
Irregularity
description
Possible cause
Solution
The auxiliary water inlet
cap is leaking.
The auxiliary water inlet
cap is worn out.
Replace it with a new one.
The auxiliary water inlet
cap is incorrectly
installed.
Install the auxiliary water inlet cap
correctly.
Irregularity
description
Possible cause
Solution
An endo-therapy
accessory does not
pass through the
instrument channel
smoothly.
An incompatible
endo-therapy accessory
is being used.
Refer to the “System chart” in the
Appendix and select a compatible
endo-therapy accessory.
Irregularity
description
Possible cause
Solution
The remote switch does
not work.
The wrong remote
switch is operated.
Operate the correct remote
switch.
The remote switch
function has been set
improperly.
Set the remote switch function
correctly as described in the video
system center’s instruction
manual.