Chapter 4 Operation
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
2.
When disconnecting the syringe or the water tube from the water pump
during examination, disconnect it directly from the luer port but leave the
auxiliary water tube itself attached.
Observation of the endoscopic image
Refer to the light source’s instruction manual for instructions on how to adjust the
brightness.
4.2
Using endo-therapy accessories
For more information on combining the endoscope with particular endo-therapy
accessories, refer to the “System chart” in the Appendix and the instruction
manuals of the accessories. Refer to the instruction manuals of the accessories
for instructions on how to operate the accessories.
• When using endo-therapy accessories, keep the distance
between the distal end of the endoscope and the mucous
membrane greater than the endoscope’s minimum visible
distance so that the endo-therapy accessory remains visible
in the endoscopic image. If the distal end of the endoscope is
placed closer than its own minimum visible distance, the
position of the accessory cannot be seen in the endoscopic
image, which could cause serious injury and/or equipment
damage. The minimum visible distance depends on the type
of endoscope being used. Refer to Section 2.3,
“Specifications”.
• When inserting or withdrawing an endo-therapy accessory,
confirm that its distal end is closed or completely retracted
into the sheath. Slowly insert or withdraw the endo-therapy
accessory straight into/from the slit of the biopsy valve.
Otherwise, the biopsy valve may be damaged and pieces of it
could fall off.
• If the insertion or withdrawal of endo-therapy accessories is
difficult, straighten the bending section as much as possible
without losing the endoscopic image. Inserting or
withdrawing endo-therapy accessories with excessive force
may damage the instrument channel or endo-therapy
accessories cause some parts to fall off and/or cause patient
injury.