Chapter 3 Preparation and Inspection
51
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
Inspection of the instrument channel
Keep your eyes away from the distal end when inserting
endo-therapy accessories. Extending the endo-therapy
accessory from the distal end could cause eye injury.
1.
Insert the endo-therapy accessory through the biopsy valve. Confirm that
the endo-therapy accessory extends smoothly from the distal end.
2.
Confirm that the endo-therapy accessory is withdrawn smoothly from the
biopsy valve.
Inspection of the auxiliary water feeding function (for
endoscopes with auxiliary water feeding only)
• Use sterile water only. Using non-sterile water may cause
patient cross-contamination and infection.
• Note that the luer port on the MAJ-855 includes a one-way
valve to prevent backflow – do not use the MAJ-855 without
this connector in place, otherwise backflow of contaminated
material may occur and equipment damage or patient injury
may result.
1.
Attach a syringe containing sterile water or the water tube from a water
pump to the luer port of the auxiliary water tube (see Figure 3.20). Feed
water and confirm that water is emitted from the auxiliary water channel at
the distal end of the insertion tube.
2.
Make sure that no water leaks at the connection between the connecting
end of the auxiliary water tube and the auxiliary water inlet.
3.
Make sure that no water leaks at the connection between the luer port of the
auxiliary water tube and the syringe or the water tube.
4.
Disconnect the water tube from the water pump or the syringe from the luer
port of the auxiliary water tube. Make sure that no water leaks from the luer
port of the auxiliary water tube and/or the distal end of the insertion tube.