Important Information — Please Read Before Use
EVIS EXERA GIF/CF/PCF TYPE 160 Series OPERATION MANUAL
• Never perform flexibility adjustment, operate the bending
section, feed air or perform suction, insert or withdraw the
endoscope’s insertion tube while the image is frozen. Never
use endo-therapy accessories while the image is frozen.
Patient injury can result.
• Regardless of the flexibility of the endoscope’s insertion tube,
never insert or withdraw it with excessive force. Otherwise,
patient injury could result.
• Do not pull the universal cord during an examination. The
endoscope connector will be pulled out from the output
socket of the light source and the endoscopic image will not
be visible.
• Do not coil the insertion tube or universal cord into a diameter
of less than 12 cm. Equipment damage can result.
• Do not touch the electrical contacts inside the electrical
connector. CCD damage may result.
• Do not apply shock to the distal end of the insertion tube,
particularly the objective lens surface at the distal end. Visual
abnormalities may result.
• Do not twist or bend the bending section with your hands.
Equipment damage may result.
• Do not squeeze the bending section forcefully. The covering
of the bending section may stretch or break and cause water
leaks.
• Turn the video system center OFF before connecting or
disconnecting the videoscope cable from the electrical
connector on the endoscope. Turn the switch ON or OFF
only when the videoscope cable is connected to both the
video system center and electrical connector on the
endoscope. Failure to do so can result in equipment damage,
including destruction of the CCD.
• The endoscope’s remote switches cannot be removed from
the control section. Pressing, pulling or twisting them with
excessive force can break the switches and/or may cause
water leaks.